OT VERIO2 METER
Report
- Report Number
- 3008382007-2016-36837
- Event Type
- Malfunction
- Date Received
- July 20, 2016
- Report Date
- July 13, 2016
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K131363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PATIENT
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
THE TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS FAILED TESTING; THE REPORTED ISSUE COULD BE CONFIRMED. THE TEST STRIPS WERE FOUND TO HAVE RESULTS ABOVE RANGE WHEN TESTED WITH CONTROL SOLUTION. ADDITIONAL TESTS WERE PERFORMED ON THE TEST STRIP VIAL AND THE DESICCANT TO CHECK THE STRUCTURAL INTEGRITY AND FOR A POSSIBLE MOISTURE ISSUE. THE STRUCTURAL INTEGRITY OF THE TEST STRIP VIAL WAS FOUND TO BE INTACT DURING A GROSS LEAK TEST. THE DESICCANT WITHIN THE SUBJECT VIAL WAS FOUND TO CONTAIN LEVELS OF MOISTURE THAT REFLECT NORMAL USE. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED
ON (B)(6) 2016, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING OBTAINING INACCURATE ERRATIC BLOOD GLUCOSE RESULTS OF ¿218, 211, 188, 196 AND 197 MG/DL¿ COMPARED TO THEIR FEELINGS AND/OR NORMAL READINGS. AT THE TIME OF TROUBLESHOOTING, THE REPORTER PERFORMED A CONTROL SOLUTION TEST AND THE RESULT FELL OUTSIDE OF THE SPECIFIED CONTROL SOLUTION RANGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 461516 | OT VERIO2 METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3882025 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |