FDA Adverse Event Malfunction Summary report: N

PCH,BSKT,DCKLING,UPBTR

MDR report key: 5806951 · Received July 19, 2016

Report

Report Number
1219602-2016-00402
Event Type
Malfunction
Date Received
July 19, 2016
Date of Event
June 22, 2016
Report Date
June 22, 2016
Manufacturer
SMITH & NEPHEW, INC.
Product Code
NBH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE INVESTIGATION NARRATIVE - ONE DUCKLING UPBITER BASKET PUNCH WAS RETURNED FOR EVALUATION. VISUAL ASSESSMENT OF THE DEVICE CONFIRMED THE REPORTED COMPLAINT OF THE LOWER JAW BREAKING OFF. EXAMINATION OF THE BREAK AREA USING A STEREO MICROSCOPE SHOWED STRIATIONS AT THE REMAINING CUTTING EDGE SURFACE WHICH IS CONSISTENT WITH THE CUTTING EDGE BEING FILED. A CUTTING EDGE IN THIS CONDITION WOULD REQUIRE THE USE OF EXCESSIVE FORCE TO CUT. THIS MANNER OF REPAIR/SHARPENING IS INCONSISTENT WITH CURRENT SMITH+NEPHEW REPAIR PRACTICES. PER THE DEVICES IFU 1060355 REV H UNDER ¿PRECAUTIONS¿ ¿AS WITH ANY SURGICAL INSTRUMENT, CAREFUL ATTENTION SHOULD BE EXERCISED TO ENSURE THAT EXCESSIVE FORCE IS NOT PLACED ON THE INSTRUMENT. EXCESSIVE FORCE CAN RESULT IN INSTRUMENT FAILURE.¿ FURTHER INVESTIGATION IS NOT WARRANTED AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A KNEE SCOPE PROCEDURE, A PIECE BROKE OFF OF THE DEVICE AND WAS RETAINED IN THE PATIENT. AN X-RAY WAS PERFORMED TO ATTEMPT TO REMOVE THE BROKEN PIECE. THE PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459552 PCH,BSKT,DCKLING,UPBTR ACCESSORIES,ARTHROSCOPIC NBH SMITH & NEPHEW, INC. 50396320

Patients

Seq Age Sex Outcome Treatment
1