HF-CABLE, MONOPOLAR, 4 M, UES-30/-40,
Report
- Report Number
- 2951238-2016-00600
- Event Type
- Malfunction
- Date Received
- July 19, 2016
- Date of Event
- June 20, 2016
- Report Date
- November 22, 2016
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- DSA
- PMA / PMN Number
- PK944201
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS NOT RETURNED TO OLYMPUS FOR EVALUATION. THE CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED; HOWEVER, USER HANDLING COULD NOT BE RULED OUT AS A CONTRIBUTING FACTOR. THE INSTRUCTION MANUAL CONTAINS SEVERAL WARNING STATEMENTS IN AN EFFORT TO PREVENT DAMAGE TO THE CABLE. ¿VISUALLY INSPECT THE CABLE AND THE PLUGS FOR IRREGULARITIES ON THE SURFACE. DO NOT USE A CABLE WITH BRITTLE OR DEFECTIVE INSULATION. REPLACE THE CABLE. IN ORDER TO PLUG OR UNPLUG THE CABLE ALWAYS PULL AT THE PLUG. NEVER PULL AT THE CABLE.¿
THE HIGH FREQUENCY (HF) CABLE WAS RETURNED TO OLYMPUS FOR EVALUATION. THE EVALUATION CONFIRMED THAT THE HF CABLE BROKEN INTO TWO PIECES AND FOUND DETACHED. THE HF CABLE COULD NOT BE TESTED DUE TO ITS RECEIVED CONDITION; HOWEVER, A MICROSCOPIC INSPECTION WAS PERFORMED ON THE RETURNED CABLE AND FOUND CHARRED STAINS INSIDE THE INTERNAL WIRES OF THE DISTAL END PLUG, WHICH IS THE PART THAT WAS CONNECTED TO THE WORKING ELEMENT. THERE WERE WHITE RESIDUES NOTED ON THE CABLE INSULATION HOUSING FROM THE PROXIMAL END PLUG. BASED ON THE INVESTIGATION FINDINGS, IT WAS DETERMINED THAT THE MOST LIKELY CAUSE OF THE REPORTED EVENT IS A SHORT ON THE INTERNAL WIRES OF THE CABLE.
OLYMPUS WAS INFORMED THAT DURING A TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE, THE HIGH FREQUENCY (HF) CABLE SPARKED AND SMOKED. THERE WAS NO INJURY TO THE USER OR PATIENT. NO EMERGENCY FIRE EVACUATION TOOK PLACE. THE PROCEDURE WAS COMPLETED USING A DIFFERENT HF CABLE. THIS IS REPORT 2 OF 4.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 456972 | HF-CABLE, MONOPOLAR, 4 M, UES-30/-40, | HF CABLE | DSA | OLYMPUS WINTER & IBE GMBH | A0393 | 342998 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |