FDA Adverse Event Malfunction Summary report: N

HF-CABLE, MONOPOLAR, 4 M, UES-30/-40,

MDR report key: 5802932 · Received July 19, 2016

Report

Report Number
2951238-2016-00598
Event Type
Malfunction
Date Received
July 19, 2016
Date of Event
June 20, 2016
Report Date
November 14, 2016
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
DSA
PMA / PMN Number
PK944201
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE DEVICE EVALUATION RESULTS AND TO UPDATE SECTIONS. THE HIGH FREQUENCY (HF) CABLE WAS RETURNED TO OLYMPUS FOR EVALUATION. THE EVALUATION WAS UNABLE TO CONFIRM THE REPORTED EVENT. OLYMPUS WAS UNABLE TO TEST THE CABLE DUE TO THE RECEIVED CONDITION. THE DISTAL END OF THE CABLE PLUG WAS DETACHED AND THE DETACHED PORTION WAS NOT RETURNED TO OLYMPUS; HOWEVER, A MICROSCOPE INSPECTION FOUND THE CABLE WITH BURNT STAINS. NO INTERNAL WIRES OF THE CABLE WERE VISIBLE. BASED ON THE INVESTIGATION FINDINGS, THE ROOT CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED AS THE DETACHED PORTION OF THE DEVICE WAS NOT RETURNED TO OLYMPUS.

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED TO OLYMPUS FOR EVALUATION. THE CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED; HOWEVER, USER HANDLING COULD NOT BE RULED OUT AS A CONTRIBUTING FACTOR. THE INSTRUCTION MANUAL CONTAINS SEVERAL WARNING STATEMENTS IN AN EFFORT TO PREVENT DAMAGE TO THE CABLE. VISUALLY INSPECT THE CABLE AND THE PLUGS FOR IRREGULARITIES ON THE SURFACE. DO NOT USE A CABLE WITH BRITTLE OR DEFECTIVE INSULATION. REPLACE THE CABLE. IN ORDER TO PLUG OR UNPLUG THE CABLE ALWAYS PULL AT THE PLUG. NEVER PULL AT THE CABLE.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT DURING A TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) PROCEDURE, THE HIGH FREQUENCY (HF) CABLE SPARKED AND SMOKED. THERE WAS NO INJURY TO THE USER OR PATIENT. NO EMERGENCY FIRE EVACUATION TOOK PLACE. THE PROCEDURE WAS COMPLETED USING A DIFFERENT HF CABLE. THIS IS REPORT 1 OF 4.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457694 HF-CABLE, MONOPOLAR, 4 M, UES-30/-40, HF CABLE DSA OLYMPUS WINTER & IBE GMBH A0393 342998

Patients

Seq Age Sex Outcome Treatment
1