FDA Adverse Event Malfunction Summary report: N

C1000 (TOUCH) THERMAL CYCLER CHASSIS

MDR report key: 5802101 · Received July 18, 2016

Report

Report Number
MW5063473
Event Type
Malfunction
Date Received
July 18, 2016
Date of Event
June 8, 2016
Report Date
July 12, 2016
Manufacturer
BIO-RAD
Product Code
OUL
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
WI, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

ON (B)(6) 2016, (B)(6) STAFF REPORTS, HEARD POPPING COMING FROM THERMAL CYCLER. UNIT STARTED SPARKING AND SMOKING. SPARKS AND FLAMES WERE NEAR THE POWER SWITCH. UNIT SERIAL NUMBER (B)(4). UNIT RETURNED TO BIO-RAD FOR REPAIR. ACCORDING TO BIO-RAD THE MALFUNCTION THAT CAUSED THE SPARKS, POPPING NOISE AND FLAMES IS DUE TO A DEFECTIVE POWER SUPPLY. BIO-RAD DID NOT ISSUE A RECALL BECAUSE OF A LESS THAN (b)(4) FAILURE RATE, AND THOUGHT THAT ISSUE IS CONTAINED TO A POWER SUPPLY ENCLOSURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
455634 C1000 (TOUCH) THERMAL CYCLER CHASSIS THERMAL CYCLER OUL BIO-RAD C1000

Patients

Seq Age Sex Outcome Treatment
1