ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2016-04022
- Event Type
- Malfunction
- Date Received
- July 15, 2016
- Date of Event
- June 27, 2016
- Report Date
- June 29, 2016
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). BATCH # K90001. THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED. THE REMAINING BLADE PORTION WAS SCRATCHED, AND WITH EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. IN ADDITION, THE CLAMP ARM WAS DETACHED FROM THE INNER TUBE, BUT FIRMLY FIXED TO THE OUTER TUBE AT THE CLAMP ARM WELD. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON GEN11 AN ALERT SCREEN WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE STOP ACTIVATING AND DISPLAY AN ALERT SCREEN IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN ALERT SCREENS, SUCH AS ¿TIGHTEN ASSEMBLY¿, ¿BLADE ERROR DETECTED¿ OR "RELAXED PRESSURE ON BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE. POSSIBLE CAUSES OF THE CLAMP ARM DETACHMENT ARE NOT CLOSING THE CLAMP WHEN SLIDING THE TORQUE WRENCH ON AND OFF; NOT CLOSING THE CLAMP WHEN INTRODUCING OR REMOVING THROUGH THE TROCAR; OR POSSIBLE ENTANGLEMENT IN FIBROUS TISSUE. THE BATCH HISTORY RECORD WAS REVIEWED AND THERE WERE NO DEFECTS, PROTOCOLS OR NCR(S) FOUND DURING THE MANUFACTURING PROCESS RELATED TO THIS COMPLAINT.
(B)(4). THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT/BATCH. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THERE WAS A TITANIUM TIP BREAKAGE. THE BREAKAGE PIECE WAS RETRIEVED BY FORCEPS AND WAS DISCARDED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.