FDA Adverse Event Malfunction Summary report: N

X SERIES

MDR report key: 5798354 · Received July 15, 2016

Report

Report Number
1220908-2016-01651
Event Type
Malfunction
Date Received
July 15, 2016
Report Date
June 27, 2016
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K112432
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION EVALUATED THE DEVICE AND THE REPORTED MALFUNCTION WAS OBSERVED AND ATTRIBUTED TO A DEFIB MONITOR FLEX CABLE THAT WAS NOT PROPERLY SEATED. THE DEVICE'S FLEX CABLE WAS REPLACED TO RESOLVE THE MALFUNCTION. NO TREND IS ASSOCIATED WITH REPORTS OF THIS TYPE.

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451019 X SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION X SERIES NA

Patients

Seq Age Sex Outcome Treatment
1