I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2016-00053
- Event Type
- Malfunction
- Date Received
- July 13, 2016
- Date of Event
- June 22, 2016
- Report Date
- September 12, 2016
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- PMA / PMN Number
- K053110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4)
(B)(4). THE INVESTIGATION WAS COMPLETED ON 08/29/2016. RETAIN PRODUCT WAS TESTED AND IS FUNCTIONING ACCORDING TO SPECIFICATION. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION.
ON 07/06/2016, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT NA, & CI RESULTS ON A PATIENT. THERE WAS NO PATIENT INFORMATION AT THE TIME OF THIS REPORT AND THE CUSTOMER STATES THEY ARE UNABLE TO PROVIDE ANY ADDITIONAL INFORMATION FOR PATIENT. THE CUSTOMER ALSO STATES THAT RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. PATIENT LOT# TEST DATE TIME NA CL PATIENT: #9, LOT# H16124, TEST DATE: (B)(6) 2016, TIME: 2025, NA: 115, CL: >140; REPEAT, H16124, (B)(6) 2016, 2033, 140, 103. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 442893 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | H16124 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |