FROVA INTUBATING INTRODUCER
Report
- Report Number
- 3002808486-2016-00699
- Event Type
- Malfunction
- Date Received
- July 13, 2016
- Date of Event
- June 14, 2016
- Report Date
- June 17, 2016
- Manufacturer
- WILLIAM COOK EUROPE
- Product Code
- LRC
- PMA / PMN Number
- E597079
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
MANUFACTURER REF# (B)(4). SUMMARY OF INVESTIGATIONAL FINDINGS: NO PRODUCT WAS RETURNED AND NO IMAGING WAS PROVIDED TO ASSIST THE INVESTIGATION AND CONSEQUENTLY IT IS VERY DIFFICULT TO DETERMINE AN EXACT REASON WHY THE "SMALL FRAGMENT OF BOUGIE SHEARED OFF AND WAS LEFT IN AIRWAY AFTER USE WITH SIZE 9 SINGLE LUMEN TUBE". IT IS NOT KNOWN, IF THE MALLINCKRODT ETT MAY HAVE HAD A SHARP EDGE ON THE DISTAL PORTION OF THE 15MM CONNECTOR WITHIN THE ETT. THE ANGLE OF INTRODUCTION OR REMOVAL OVER THE FROVA HAS THEN POTENTIALLY CAUSED A SHAVING OF FROVA MATERIAL TO OCCUR. THE MATERIAL IS RADIOPAQUE EXTRUDED POLYETHYLENE AND THE MATERIAL CHARACTERISTICS DO NOT SUGGEST THAT FLAKING WOULD OCCUR, BUT IF IT WAS TO ENCOUNTER A ROUGH EDGE IT MAY BE POSSIBLE. THIS IS THE FIRST COMPLAINT RECEIVED REPORTING THAT A FRAGMENT SHEARED OFF FROM THE FROVA INTRODUCER, WHEN USED WITH A SINGLE LUMEN TUBE. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS.
(B)(4). LOT #: UNKNOWN. EXPIRATION DATE: UNKNOWN AS LOT# IS UNKNOWN. MFR DATE UNKNOWN. INVESTIGATION IS STILL IN PROGRESS.
DESCRIPTION OF EVENT ACCORDING TO COMPLAINANT: SMALL FRAGMENT OF BOUGIE SHEARED OFF AND LEFT IN AIRWAY AFTER USE WITH SIZE 9 SINGLE LUMEN. NOTICE ON ROUTINE BRONCHOSCOPY IMMEDIATELY AFTER INTUBATION AND REMOVED BY SUCTION. ADDITIONAL INFORMATION RECEIVED FROM PHYSICIAN: AS PART OF A VIDEO ASSISTED THORACOSCOPIC SURGERY (VATS) FOR LOBECTOMY, THE THORACIC SURGEON PERFORMED A FLEXIBLE BRONCHOSCOPY VIA A SIZE 9 ENDOTRACHEAL TUBE (MALLINCKRODT HI CONTOUR, COVIDIEN IRELAND LIMITED). TO EVERYONE'S SURPRISE IT SHOWED A PIECE OF BRIGHT BLUE PLASTIC IN THE TRACHEA WHICH WAS SUCCESSFULLY REMOVED WITH SUCTION. THE ETT HAD BEEN SITED USING A FROVA INTUBATING CATHETER (WILLIAM COOK EUROPE APS) DUE TO A CORMACK-LEHANE GRADE 3 VIEW. ON CLOSER INSPECTION OF THE BOUGIE, AN EXTREMELY THIN SLICE OF ITS BLUE COATING HAD BEEN SHAVED OFF AT THE OUTSIDE EDGE OF THE ANGLED TIP. THE INTUBATION WAS ATRAUMATIC AND THE REMOVAL OF THE BOUGIE DID NOT REQUIRE MORE FORCE THAN USUAL. PATIENT OUTCOME: ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE
DESCRIPTION OF EVENT ACCORDING TO COMPLAINANT: SMALL FRAGMENT OF BOUGIE SHEARED OFF AND LEFT IN AIRWAY AFTER USE WITH SIZE 9 SINGLE LUMEN. NOTICE ON ROUTINE BRONCHOSCOPY IMMEDIATELY AFTER INTUBATION AND REMOVED BY SUCTION. ADDITIONAL INFORMATION RECEIVED FROM PHYSICIAN: AS PART OF A VIDEO ASSISTED THORACOSCOPIC SURGERY (VATS) FOR LOBECTOMY, THE THORACIC SURGEON PERFORMED A FLEXIBLE BRONCHOSCOPY VIA A SIZE 9 ENDOTRACHEAL TUBE (MALLINCKRODT HI CONTOUR, COVIDIEN (B)(4)). TO EVERYONE¿S SURPRISE IT SHOWED A PIECE OF BRIGHT BLUE PLASTIC IN THE TRACHEA WHICH WAS SUCCESSFULLY REMOVED WITH SUCTION. THE ETT HAD BEEN SITED USING A FROVA INTUBATING CATHETER (WILLIAM COOK (B)(4)) DUE TO A CORMACK-LEHANE GRADE 3 VIEW. ON CLOSER INSPECTION OF THE BOUGIE, AN EXTREMELY THIN SLICE OF ITS BLUE COATING HAD BEEN SHAVED OFF AT THE OUTSIDE EDGE OF THE ANGLED TIP. THE INTUBATION WAS ATRAUMATIC AND THE REMOVAL OF THE BOUGIE DID NOT REQUIRE MORE FORCE THAN USUAL. PATIENT OUTCOME: ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 443201 | FROVA INTUBATING INTRODUCER | LRC INSTRUMENT, ENT MANUAL SURGICAL | LRC | WILLIAM COOK EUROPE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |