DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2016-46067
- Event Type
- Malfunction
- Date Received
- July 12, 2016
- Date of Event
- June 14, 2016
- Report Date
- June 14, 2016
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- UDI-DI
- 00386270000019
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
PATIENT'S MOTHER CONTACTED DEXCOM ON (B)(6) 2016 TO REPORT CONTINUOUS GLUCOSE MONITORING (CGM) INACCURACIES COMPARED TO BLOOD GLUCOSE (BG) METER THAT OCCURRED ON (B)(6) 2016. SENSOR WAS INSERTED ON (B)(6) 2016. AT THE TIME OF CONTACT, NO INJURY OR MEDICAL INTERVENTION WAS REPORTED. THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF DEVICE EVALUATION. THE RECEIVER DATA LOG WAS REVIEWED ON 06/30/2016. THE COMPLAINT OF INACCURATE CGM VALUES WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.
SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT SENSOR WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER (PART NUMBER STT-GF-001/SERIAL NUMBER (B)(4)/LOT NUMBER 5210262) WAS RETURNED ON 07/07/2016. THE RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND FAILED TESTING. THE REPORTED EVENT OF INACCURACIES WAS NOT CONFIRMED BECAUSE INFORMATION ABOUT INACCURACIES CANNOT BE OBTAINED FROM THE TRANSMITTER. A ROOT CAUSE COULD NOT BE DETERMINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 439201 | DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9500-27 | 5211349 | 00386270000019 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |