FDA Adverse Event
Other
Summary report: N
BD PARADIGM LINK BLOOD GLUCOSE MONITOR
MDR report key: 578804
·
Received February 22, 2005
Report
- Report Number
- 2243072-2005-00043
- Event Type
- Other
- Date Received
- February 22, 2005
- Date of Event
- January 21, 2005
- Report Date
- February 18, 2005
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER CALLED TO REPORT GETTING DIFFERENT READINGS WHEN USING THEIR FAMILY MEMBER'S PARADIGM LINK METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVERAGE READING (235) AS REF.POINT VS. BD READINGS OF 400, 178, 128 YIELDED DATA POINTS IN THE C ZONE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD PARADIGM LINK BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 4064222 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |