FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 578804 · Received February 22, 2005

Report

Report Number
2243072-2005-00043
Event Type
Other
Date Received
February 22, 2005
Date of Event
January 21, 2005
Report Date
February 18, 2005
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT GETTING DIFFERENT READINGS WHEN USING THEIR FAMILY MEMBER'S PARADIGM LINK METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVERAGE READING (235) AS REF.POINT VS. BD READINGS OF 400, 178, 128 YIELDED DATA POINTS IN THE C ZONE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 4064222

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other