FDA Adverse Event Injury Summary report: N

ZYSTON STRAIGHT INLINE INSERTER

MDR report key: 5786167 · Received July 12, 2016

Report

Report Number
3004485144-2016-00144
Event Type
Injury
Date Received
July 12, 2016
Report Date
March 16, 2017
Manufacturer
BIOMET SPINE - BROOMFIELD
Product Code
MAX
PMA / PMN Number
PK112014
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

WITHOUT A PRODUCT RETURN, NO PRODUCT EVALUATION IS ABLE TO BE CONDUCTED. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. IF ADDITIONAL INFORMATION IS OBTAINED THAT ADDS VALUE TO THE RELEVANT CONTENT OF THIS REPORT AND/OR A CONCLUSION CAN BE DRAWN, A FOLLOW-UP REPORT WILL BE SENT. REPORT TWO OF THREE FOR THE SAME EVENT; SEE ALSO 3004485144-2016-00143 AND 00145.

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. A FOLLOW UP REPORT WILL BE SENT UPON COMPLETION OF THE DEVICE EVALUATION. REPORT TWO OF THREE FOR THE SAME EVENT; SEE ALSO 3004485144-2016-00143-1 AND 00145-1.

Additional Manufacturer Narrative · 1

THE RETURNED INSERTER WAS EXAMINED. THERE WERE NO SIGNS OF COSMETIC DAMAGE OR DEFORMITIES. THERE IS NO EVIDENCE THAT THE METAL FRAGMENTS WERE PRODUCED FROM THIS INSERTER. A REVIEW OF THE MANUFACTURING RECORDS DID NOT IDENTIFY ANY ISSUES WHICH WOULD HAVE CONTRIBUTED TO THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE THE SURGEON WAS IMPACTING A CAGE WITH ZYSTON STRAIGHT INSERTER STRAIGHT HANDLE AT MIS-TLIF OPERATION; SOMETHING LIKE SMALL METAL PARTICLES/POWDER WERE SEEN FROM THE INSERTER HANDLE OR SHAFT. METAL PARTICLES FROM THE BLACK 1/4 SQUARE T-HANDLE UPON IMPACTION, WAS ALSO OBSERVED. IT IS UNKNOWN IF THE SMALL PARTICLES/POWDER FELL INTO THE PATIENT. THE SURGERY WAS COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
441205 ZYSTON STRAIGHT INLINE INSERTER INSERTER MAX BIOMET SPINE - BROOMFIELD N/A 038281

Patients

Seq Age Sex Outcome Treatment
1 Other