ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2016-15694
- Event Type
- Malfunction
- Date Received
- July 12, 2016
- Report Date
- June 22, 2016
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
FOLLOW-UP #1: DATE OF SUBMISSION 09/22/2016 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/27/2016 WITH THE FOLLOWING FINDINGS:THE BATTERY COMPARTMENT IS CRACKED BELOW THE GRIP PAD, THE PUMP POWERS UP WITH AUDITORY AND VIBRATION TO A BLANK SCREEN. REPORTED ¿BLANK SCREEN¿ COMPLAINT IS DUPLICATED. THE PUMP¿S COVER WAS REMOVED, THE DISPLAY SCREEN WAS FOUND CRACKED. A TEST DISPLAY WAS MOUNTED, BUT THE PUMP STILL POWERS UP WITH AUDITORY AND VIBRATION TO A BLANK SCREEN. TEST CODE DOWNLOADER TOOL APPLICATION V 1.0.0 WAS USED TO UPLOAD TEST CODE V 1.1.12 TO MASTER/SLAVE/PERIPH PROCESSORS AS PER (6000975), THE UPLOAD WAS SUCCESSFUL. THE PUMP POWERS UP AND DISPLAY JUST ¿MSP¿ , NOT ¿MSP2¿, (2) IS THE EEPROM COMMUNICATING PROPERLY HENCE THE DIAGNOSIS THAT THE FAILURE IS IN THIS AREA. PCB INSPECTION FOUND LOWER EEPROM U22 CRACKED. EEPROM FAILURE IS CONCLUDED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 441469 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR |