INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Report
- Report Number
- 9611451-2016-00330
- Event Type
- Malfunction
- Date Received
- July 11, 2016
- Date of Event
- June 6, 2016
- Report Date
- June 13, 2016
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K103767
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT WAS RETURNED TO FPH IN (B)(4) FOR EVALUATION, WHERE IT WAS VISUALLY INSPECTED AND PRESSURE TESTED. RESULTS: VISUAL INSPECTION OF THE SUBJECT DEVICE REVEALED A CRACK ON THE EXPIRATORY LIMB PATIENT END CONNECTOR COLLAR. NO VISIBLE DAMAGE WAS OBSERVED ON THE BREATHING CIRCUIT TUBING. THE PRESSURE TEST RESULT WAS OUTSIDE SPECIFICATION. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 160119. CONCLUSION: BASED ON THE NATURE OF THE CRACKING AND PREVIOUS INVESTIGATIONS INTO THIS TYPE OF FAILURE, THE CRACKING IS MOST LIKELY DUE TO THE CONNECTORS COMING INTO CONTACT WITH A CLEANING CHEMICAL, RESULTING IN ENVIRONMENTAL STRESS CRACKING. ALL RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE AND FLOW TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. THE HOSPITAL REPORTED THAT THE DAMAGE OCCURRED AFTER 6 DAYS OF USE, WHICH SUGGESTS THAT THE COMPLAINT CIRCUIT BECAME DAMAGED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE."; "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."; "SET APPROPRIATE VENTILATOR ALARMS." THE USER INSTRUCTIONS ALSO STATE THE FOLLOWING: - "DO NOT SOAK, WASH, STERILIZE OR REUSE THIS PRODUCT. AVOID CONTACT WITH CHEMICALS, CLEANING AGENTS OR HAND SANITISERS."
A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT HAD A CRACK ON THE EXPIRATORY LIMB COLLAR AFTER 6 DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 435559 | INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT265 | 2100019372 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |