FDA Adverse Event Malfunction Summary report: N

INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 5784441 · Received July 11, 2016

Report

Report Number
9611451-2016-00330
Event Type
Malfunction
Date Received
July 11, 2016
Date of Event
June 6, 2016
Report Date
June 13, 2016
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K103767
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT WAS RETURNED TO FPH IN (B)(4) FOR EVALUATION, WHERE IT WAS VISUALLY INSPECTED AND PRESSURE TESTED. RESULTS: VISUAL INSPECTION OF THE SUBJECT DEVICE REVEALED A CRACK ON THE EXPIRATORY LIMB PATIENT END CONNECTOR COLLAR. NO VISIBLE DAMAGE WAS OBSERVED ON THE BREATHING CIRCUIT TUBING. THE PRESSURE TEST RESULT WAS OUTSIDE SPECIFICATION. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 160119. CONCLUSION: BASED ON THE NATURE OF THE CRACKING AND PREVIOUS INVESTIGATIONS INTO THIS TYPE OF FAILURE, THE CRACKING IS MOST LIKELY DUE TO THE CONNECTORS COMING INTO CONTACT WITH A CLEANING CHEMICAL, RESULTING IN ENVIRONMENTAL STRESS CRACKING. ALL RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE AND FLOW TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. THE HOSPITAL REPORTED THAT THE DAMAGE OCCURRED AFTER 6 DAYS OF USE, WHICH SUGGESTS THAT THE COMPLAINT CIRCUIT BECAME DAMAGED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE."; "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."; "SET APPROPRIATE VENTILATOR ALARMS." THE USER INSTRUCTIONS ALSO STATE THE FOLLOWING: - "DO NOT SOAK, WASH, STERILIZE OR REUSE THIS PRODUCT. AVOID CONTACT WITH CHEMICALS, CLEANING AGENTS OR HAND SANITISERS."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT HAD A CRACK ON THE EXPIRATORY LIMB COLLAR AFTER 6 DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
435559 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT265 2100019372

Patients

Seq Age Sex Outcome Treatment
1