FDA Adverse Event Malfunction Summary report: N

IPC® HANDPIECE ATTACHMENT - INDIGO¿ DRILL

MDR report key: 5783966 · Received July 11, 2016

Report

Report Number
1625507-2016-00030
Event Type
Malfunction
Date Received
July 11, 2016
Date of Event
June 15, 2016
Report Date
June 15, 2016
Manufacturer
MDT POWERED SURGICAL SOLUTIONS
Product Code
HBE
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE 1845000 (INDIGO DRILL) <(>&<)> 1845020 (INDIGO ANGLED ATTACHMENT): THE PRODUCT ANALYSIS INDICATES THAT THE REPORTED OVERHEATING WAS CONFIRMED. THE ONE-MINUTE PRE-REPAIR TEMPERATURE TEST RESULTS WERE 129 DEGREES F AND 133 DEGREES F. THE DRILL FAILED TESTS. NO REPAIRS WERE PERFORMED. (B)(4). CONCLUSION: USE ERROR CAUSED OR CONTRIBUTED TO EVENT .

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCT: 1845000 ¿ DRILL, INDIGO HIGH SPEED OTOLOGIC; SERIAL NUMBER ¿ (B)(4); LOT NUMBER - 209989479; MANUFACTURE DATE ¿ AUGUST 13, 2015; 510K - K081475. INITIAL INSPECTION PERFORMED BY MEDTRONIC (B)(4) INDICATES THAT THE SERVICE TECHNICIAN WAS UNABLE TO LOCK THE ATTACHMENT WITH THE BUR AND THE BUR LOCK WAS NOT TURNING. THE ATTACHMENT IS TO BE FORWARDED TO THE U.S. FOR ANALYSIS. 1845000 (DRILL): INITIAL INSPECTION PERFORMED BY MEDTRONIC (B)(4) INDICATES THAT THE DRILL WAS WORKING WELL. THE REPORTED OVERHEATING COULD NOT BE CONFIRMED. THE ONE-MINUTE PRE-REPAIR TEMPERATURE TEST RESULTS WERE AS FOLLOWS: 88.2 DEGREES F AT T1 AND 85.5 DEGREES F AT T2.

Description of Event or Problem · 1

THE AVAILABLE INFORMATION INDICATES THAT PRIOR TO USE, THE DRILL AND ATTACHMENT OVERHEATED RAPIDLY. THERE WAS NO INJURY REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
435424 IPC® HANDPIECE ATTACHMENT - INDIGO¿ DRILL DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MDT POWERED SURGICAL SOLUTIONS 1845020 210944762

Patients

Seq Age Sex Outcome Treatment
1