FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 5783927 · Received July 11, 2016

Report

Report Number
2245578-2016-00047
Event Type
Malfunction
Date Received
July 11, 2016
Date of Event
June 23, 2016
Report Date
September 12, 2016
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
PMA / PMN Number
K053110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION WAS COMPLETED ON 08/29/2016. RETAIN PRODUCT WAS TESTED AND IS FUNCTIONING ACCORDING TO SPECIFICATION. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION.

Description of Event or Problem · 1

ON 07/06/2016, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT NA, & CI RESULTS ON A PATIENT. THERE WAS NO PATIENT INFORMATION AT THE TIME OF THIS REPORT AND THE CUSTOMER STATES THEY ARE UNABLE TO PROVIDE ANY ADDITIONAL INFORMATION FOR PATIENT. THE CUSTOMER ALSO STATES THAT RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. PATIENT: #3, LOT# H16124, TEST DATE: (B)(6) 2016, TIME: 1055, NA: 105, CL: >140; REPEAT, H16124, (B)(6) 2016, 1112, 138, 101. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.

Description of Event or Problem · 1

NA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
438288 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JGS ABBOTT POINT OF CARE NA H16124

Patients

Seq Age Sex Outcome Treatment
1