FDA Adverse Event Malfunction Summary report: N

E SERIES DEFIBRILLATOR

MDR report key: 5783745 · Received July 11, 2016

Report

Report Number
1220908-2016-01630
Event Type
Malfunction
Date Received
July 11, 2016
Report Date
June 23, 2016
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K042007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY INTERCONNECT FLEX CABLE. THE FLEX CABLE WAS REPLACED TO RESOLVE THE MALFUNCTION. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE PRODUCT FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY BLANKED OUT. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
437377 E SERIES DEFIBRILLATOR E SERIES MKJ ZOLL MEDICAL CORPORATION E SERIES NA

Patients

Seq Age Sex Outcome Treatment
1