FDA Adverse Event Injury Summary report: N

COCR BALL HEAD 12/14 Ø 28 SIZE M 0

MDR report key: 5778952 · Received July 8, 2016

Report

Report Number
3005180920-2016-00338
Event Type
Injury
Date Received
July 8, 2016
Date of Event
June 8, 2016
Report Date
July 8, 2016
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
PMA / PMN Number
K072857
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 04 JULY 2016 THE (B)(4) PERFORMED A CLINICAL EVALUATION AND COMMENTED AS FOLLOWS: DAY 1 DISLOCATION IN THA IS A POSSIBLE ADVERSE EVENT, WHICH NORMALLY COMES FROM DIFFICULTIES IN ACHIEVING EITHER A SATISFACTORY RELATIVE POSITION OF THE COMPONENTS OR INSUFFICIENT SOFT-TISSUE TENSION. IN THIS CASE, NO INFORMATION IS AVAILABLE IN TERMS OF X-RAYS OR FURTHER INTRA-OPERATIVE DETAILS. REPLACEMENT OF THE HEAD WITH A LONGER-NECK ONE SUGGESTS THAT INSUFFICIENT SOFT-TISSUE TENSION HAD BEEN ACHIEVED AT PRIMARY OPERATION. WHETHER THIS IS DUE TO STEM SUBSIDENCE OR CUP MIGRATION OR TO A PROBLEM IN INTRA-OPERATIVE TENSION ASSESSMENT IS UNKNOWN, BUT IS VERY UNLIKELY TO BE ATTRIBUTED TO A DEFECTIVE DEVICE. BATCH REVIEW PERFORMED ON 05 JULY 2016. LOT 151904: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 24 AUGUST 2015. EXPIRATION DATE: 2020-06-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ON 08 JULY 2016 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION SUBMITTED IN THIS INITIAL REPORT. ON THE SAME DATE THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.

Description of Event or Problem · 1

AFTER THE SURGERY, THE SURGEON FOUND DISLOCATION. HE IMMEDIATELY PERFORMED REVISION SURGERY AND REPLACED THE FEMORAL HEAD WITH MORE LARGER SIZE. EXPLANTS AND X-RAYS ARE NOT AVAILABLE FOR INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
433300 COCR BALL HEAD 12/14 Ø 28 SIZE M 0 METALLIC BALL HEAD JDI MEDACTA INTERNATIONAL SA 151904

Patients

Seq Age Sex Outcome Treatment
1 81 YR Required Intervention