PRODIGY VOICE
Report
- Report Number
- 3005862821-2016-00034
- Event Type
- Injury
- Date Received
- July 8, 2016
- Date of Event
- April 15, 2016
- Report Date
- June 16, 2016
- Manufacturer
- OK BIOTECH CO., LTD.
- Product Code
- NBW
- PMA / PMN Number
- K073118
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
SUSPECTED DEVICE EVALUATED BY OK BIOTECH AND CALCULATED THAT THE METER OPERATED WITHIN SPECIFICATIONS. THE STANDBY CURRENT TEST IS 12.7. THE CRITERIA IS <55 PASS. METER SETTING, AUDIO AND ALL BUTTONS FUNCTION ARE OK. TESTED THE SUSPECTED METER WITH IN HOUSE CONTROL SOLUTION AND RETAIN STRIPS (STRIP LOT NUMBER: D150825-1). THE CONTROL SOLUTION TESTS FOR LEVEL LOW ARE 68/66 MG/DL,FOR LEVEL HIGH ARE 276/268 MG/DL. THE CONTROL SOLUTION RANGES ARE: LEVEL LOW 25~70 MG/DL; LEVEL HIGH 210~310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS. RETURNED STRIPS TEST (STRIP LOT NUMBER:D150825-1). THE CONTROL SOLUTION TESTS FOR LEVEL LOW WERE 67/66 MG/DL, FOR LEVEL HIGH WERE 283/285 MG/DL. THE CONTROL SOLUTION RANGES ARE: LEVEL LOW 25~70 MG/DL; LEVEL HIGH 210~310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS.
OUR IMPORTER, (B)(4) RECEIVED A CALL ON (B)(6) 2016 REPORTING A MEDICAL INTERVENTION THAT OCCURRED ON (B)(6) 2016 AT 2:00PM. PATIENT'S REHAB SPECIALIST CALLED IN STATING THAT PATIENT'S BLOOD GLUCOSE WAS LOW. PARAMEDICS WERE CALLED 1 HOUR AFTER TESTING WITH THE (B)(4) METER. THE READING ON THE (B)(4) METER AT THE TIME OF THE EVENT WAS UNKNOWN. PARAMEDICS WERE CALLED 1 HOUR AFTER TESTING WITH THE (B)(4) METER. PARAMEDICS ARRIVED APPROXIMATELY 7 MINUTES AFTER BEING CALLED. PATIENT'S NORMAL BLOOD GLUCOSE READING AROUND THE TIME OF THE EVENT IS 175MG/DL. REHAB SPECIALIST STATED THAT PARAMEDICS TESTED PATIENT'S GLUCOSE WITH THE PARAMEDIC'S METER AND IT WAS 20MG/DL LOWER THAN THE (B)(4) METER. PARAMEDICS TRANSPORTED PATIENT TO ER. UPON ARRIVAL REHAB SPECIALIST STATED THAT PATIENT'S GLUCOSE LEVEL WAS 20MG/DL LOWER THAN THE (B)(4) METER. (B)(4) SENT REPLACEMENT AND PREPAID ENVELOPE REQUESTING RETURN OF SUSPECT DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 431636 | PRODIGY VOICE | BLOOD GLUCOSE MONITORING DEVICE | NBW | OK BIOTECH CO., LTD. | 51900 | D150825-1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Required Intervention | CALCIUM |