FDA Adverse Event Malfunction Summary report: N

AUTOCAT2 WAVE FRENCH

MDR report key: 5776642 · Received July 7, 2016

Report

Report Number
1219856-2016-00151
Event Type
Malfunction
Date Received
July 7, 2016
Date of Event
June 10, 2016
Report Date
June 13, 2016
Manufacturer
ARROW INTERNATIONAL INC.
Product Code
DSP
PMA / PMN Number
K060309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

QN#(B)(4). THE PCS ASSEMBLY (P/N 96-3006-001W, S/N (B)(4)) WAS RETURNED FOR EVALUATION. VISUAL INSPECTION OF THE PCS ASSEMBLY WAS PERFORMED AND NO ABNORMALITIES WERE NOTED. THE PCS ASSEMBLY IN QUESTION WAS INSTALLED INTO KNOWN GOOD AUTOCAT2W AND PERFORMED FUNCTIONAL TESTING. THE KNOWN GOOD AUTOCAT2W WITH THE PCS ASSEMBLY IN QUESTION INSTALLED FAILED THE FUNCTIONAL TEST. THE PUMP ALARMED "HELIUM LOSS (2)" APPROXIMATELY 15 MINUTES AFTER INITIATED PUMPING. THE PCS ASSEMBLY WAS THEN REMOVED FROM THE PUMP. THE PCS ASSEMBLY IN QUESTION WAS INSTALLED ONTO THE LEAK TESTER AND FAILED LEAK TESTING. A SMALL LEAK WAS DETECTED TO BE COMING FROM THE VENT PORT V1. THE VENT VALVE V1 WAS POWERED WITH 12 VDC USING EXTERNAL POWER SUPPLY AND CLICKING SOUND WAS NOTED. THE EXTERNAL POWER SUPPLY WAS THEN SWITCHED ON/OFF MULTIPLE TIMES WITH TRIGGER EACH TIME. VISUAL INSPECTION OF THE PCS ASSEMBLY INTERNAL HARDWARE WAS PERFORMED AND NO ABNORMALITIES WERE NOTED. A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE LOT NUMBER/SERIAL NUMBER WITH NO RELEVANT FINDINGS. THE DEVICE PASSED ALL MANUFACTURING SPECIFICATIONS PRIOR TO RELEASE. SEE OTHER REMARKS SECTION. OTHER REMARKS: CONCLUSION: THE REPORTED COMPLAINT OF "ALARM, HELIUM LOSS" IS CONFIRMED. THE REPORTED ALARM WAS REPLICATED DURING THE FUNCTIONAL TEST. THE PUMP ALARMED HELIUM LOSS (2) APPROXIMATELY 15 MINUTES AFTER INITIATED PUMPING. A SMALL LEAK WAS DETECTED TO BE COMING FROM THE VENT PORT, WHICH CAUSED THE ALARM. THE CAUSE OF VENT VALVE MALFUNCTION IS UNDETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED VIA A FIELD SERVICE REPORT: 061016M. SYMPTOM: HELIUM LEAK. FINDINGS / ACTION TAKEN: PCS ASSEMBLY REPLACED. PUMP CHECKED: PASS. SOFTWARE: 2.24FR. FCN: 1416. REASON FOR SERVICE: OTHER REASON. RESULTS: OPERATIONAL. PER THE TECHNICIAN SAV CONTRE PULSION IN (B)(6): "THE ERROR WAS REPORTED BY THE USER DEPARTMENT WHEN USED ON PATIENTS WITHOUT MORE INFORMATION FROM PATIENT'S CONDITION. THE PUMP WAS REPLACED BY ANOTHER PUMP. THE HOSPITAL DID NOT FILE COMPLAINTS".

Description of Event or Problem · 1

IT WAS REPORTED VIA A FIELD SERVICE REPORT: (B)(4). SYMPTOM: HELIUM LEAK. FINDINGS / ACTION TAKEN: PCS ASSEMBLY REPLACED. PUMP CHECKED: PASS. SOFTWARE: 2.24FR. FCN: 1416. REASON FOR SERVICE: OTHER REASON. RESULTS: OPERATIONAL. PER THE TECHNICIEN (B)(6): "THE ERROR WAS REPORTED BY THE USER DEPARTMENT WHEN USED ON PATIENTS WITHOUT MORE INFORMATION FROM PATIENT'S CONDITION. THE PUMP WAS REPLACED BY ANOTHER PUMP. THE HOSPITAL DID NOT FILE COMPLAINTS".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
429409 AUTOCAT2 WAVE FRENCH INTRA-AORTIC BALLOON PUMP DSP ARROW INTERNATIONAL INC. 0001052227

Patients

Seq Age Sex Outcome Treatment
1