AUTOCAT2 WAVE FRENCH
Report
- Report Number
- 1219856-2016-00151
- Event Type
- Malfunction
- Date Received
- July 7, 2016
- Date of Event
- June 10, 2016
- Report Date
- June 13, 2016
- Manufacturer
- ARROW INTERNATIONAL INC.
- Product Code
- DSP
- PMA / PMN Number
- K060309
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4).
QN#(B)(4). THE PCS ASSEMBLY (P/N 96-3006-001W, S/N (B)(4)) WAS RETURNED FOR EVALUATION. VISUAL INSPECTION OF THE PCS ASSEMBLY WAS PERFORMED AND NO ABNORMALITIES WERE NOTED. THE PCS ASSEMBLY IN QUESTION WAS INSTALLED INTO KNOWN GOOD AUTOCAT2W AND PERFORMED FUNCTIONAL TESTING. THE KNOWN GOOD AUTOCAT2W WITH THE PCS ASSEMBLY IN QUESTION INSTALLED FAILED THE FUNCTIONAL TEST. THE PUMP ALARMED "HELIUM LOSS (2)" APPROXIMATELY 15 MINUTES AFTER INITIATED PUMPING. THE PCS ASSEMBLY WAS THEN REMOVED FROM THE PUMP. THE PCS ASSEMBLY IN QUESTION WAS INSTALLED ONTO THE LEAK TESTER AND FAILED LEAK TESTING. A SMALL LEAK WAS DETECTED TO BE COMING FROM THE VENT PORT V1. THE VENT VALVE V1 WAS POWERED WITH 12 VDC USING EXTERNAL POWER SUPPLY AND CLICKING SOUND WAS NOTED. THE EXTERNAL POWER SUPPLY WAS THEN SWITCHED ON/OFF MULTIPLE TIMES WITH TRIGGER EACH TIME. VISUAL INSPECTION OF THE PCS ASSEMBLY INTERNAL HARDWARE WAS PERFORMED AND NO ABNORMALITIES WERE NOTED. A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE LOT NUMBER/SERIAL NUMBER WITH NO RELEVANT FINDINGS. THE DEVICE PASSED ALL MANUFACTURING SPECIFICATIONS PRIOR TO RELEASE. SEE OTHER REMARKS SECTION. OTHER REMARKS: CONCLUSION: THE REPORTED COMPLAINT OF "ALARM, HELIUM LOSS" IS CONFIRMED. THE REPORTED ALARM WAS REPLICATED DURING THE FUNCTIONAL TEST. THE PUMP ALARMED HELIUM LOSS (2) APPROXIMATELY 15 MINUTES AFTER INITIATED PUMPING. A SMALL LEAK WAS DETECTED TO BE COMING FROM THE VENT PORT, WHICH CAUSED THE ALARM. THE CAUSE OF VENT VALVE MALFUNCTION IS UNDETERMINED.
IT WAS REPORTED VIA A FIELD SERVICE REPORT: 061016M. SYMPTOM: HELIUM LEAK. FINDINGS / ACTION TAKEN: PCS ASSEMBLY REPLACED. PUMP CHECKED: PASS. SOFTWARE: 2.24FR. FCN: 1416. REASON FOR SERVICE: OTHER REASON. RESULTS: OPERATIONAL. PER THE TECHNICIAN SAV CONTRE PULSION IN (B)(6): "THE ERROR WAS REPORTED BY THE USER DEPARTMENT WHEN USED ON PATIENTS WITHOUT MORE INFORMATION FROM PATIENT'S CONDITION. THE PUMP WAS REPLACED BY ANOTHER PUMP. THE HOSPITAL DID NOT FILE COMPLAINTS".
IT WAS REPORTED VIA A FIELD SERVICE REPORT: (B)(4). SYMPTOM: HELIUM LEAK. FINDINGS / ACTION TAKEN: PCS ASSEMBLY REPLACED. PUMP CHECKED: PASS. SOFTWARE: 2.24FR. FCN: 1416. REASON FOR SERVICE: OTHER REASON. RESULTS: OPERATIONAL. PER THE TECHNICIEN (B)(6): "THE ERROR WAS REPORTED BY THE USER DEPARTMENT WHEN USED ON PATIENTS WITHOUT MORE INFORMATION FROM PATIENT'S CONDITION. THE PUMP WAS REPLACED BY ANOTHER PUMP. THE HOSPITAL DID NOT FILE COMPLAINTS".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 429409 | AUTOCAT2 WAVE FRENCH | INTRA-AORTIC BALLOON PUMP | DSP | ARROW INTERNATIONAL INC. | 0001052227 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |