AUTOCAT2 WAVE
Report
- Report Number
- 1219856-2016-00150
- Event Type
- Malfunction
- Date Received
- July 7, 2016
- Date of Event
- June 9, 2016
- Report Date
- June 9, 2016
- Manufacturer
- ARROW INTERNATIONAL INC.
- Product Code
- DSP
- PMA / PMN Number
- K060309
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
QN#(B)(4). THE PCS ASSEMBLY (P/N 96-3006-001, S/N (B)(4)) WAS RETURNED FOR EVALUATION. VISUAL INSPECTION OF THE PCS ASSEMBLY WAS PERFORMED AND NO ABNORMALITY WAS FOUND. THE PCS ASSEMBLY IN QUESTION WAS INSTALLED INTO KNOWN GOOD AUTOCAT2W AND PERFORMED FUNCTIONAL TESTING. THE KNOWN GOOD AUTOCAT2W WITH THE PCS ASSEMBLY IN QUESTION INSTALLED PASSED THE FUNCTIONAL TEST. THE PURGE CYCLE WAS PERFORMED MULTIPLE TIMES WITH NO ALARMS OR ERRORS. THE PUMP WAS THEN LEFT TO RUN FOR OVER SIX HOURS WITHOUT ANY ISSUES. THE PCS ASSEMBLY IN QUESTION WAS INSTALLED ONTO THE MDT-50 LEAK TESTER AND PASSED LEAK TESTING. VISUAL INSPECTION OF THE PCS ASSEMBLY INTERNAL HARDWARE WAS PERFORMED AND NO ABNORMALITY WAS FOUND. PER SERVICE HISTORY REPORTED, THE PCS ASSEMBLY WAS REPLACED ON NOVEMBER 22, 2005. A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE IABP SERIAL/ LOT NUMBER WITH NO RELEVANT FINDINGS. THE DEVICE PASSED ALL MANUFACTURING SPECIFICATIONS PRIOR TO RELEASE. OTHER REMARKS: CONCLUSION: THE REPORTED PROBLEM OF PURGE FAILURE ALARM IS NOT CONFIRMED. THE REPORTED PROBLEM COULD NOT BE REPRODUCED DURING THE FUNCTIONAL TEST. THE PCS ASSEMBLY PASSED FUNCTIONAL TESTING. THE CAUSE OF THE REPORTED COMPLAINT IS UNDETERMINED.
(B)(4).
IT WAS REPORTED VIA A FIELD SERVICE REPORT - L610990. PUMP HAD PURGE FAILURE ALARM WHEN FIRST PUT ON PATIENT. PUMP WAS SWITCHED OUT QUICKLY WITHOUT DELAY OR IMPACT TO PATIENT. FINDINGS/ACTIONS TAKEN: CONFIRMED- PCS ASSEMBLY REPLACED. PARTS SENT TO AND REPLACED BY HOSPITAL BIOMED. DEFECTIVE PCS ASSY. TO BE RETURNED ON (B)(4). FCN LEVEL: 1416, SOFTWARE LEVEL: 2.24.
IT WAS REPORTED VIA A FIELD SERVICE REPORT - (B)(4). PUMP HAD PURGE FAILURE ALARM WHEN FIRST PUT ON PATIENT. PUMP WAS SWITCHED OUT QUICKLY WITHOUT DELAY OR IMPACT TO PATIENT.FINDINGS/ACTIONS TAKEN: CONFIRMED- PCS ASSEMBLY REPLACED. PARTS SENT TO AND REPLACED BY HOSPITAL BIOMED. DEFECTIVE PCS ASSY. TO BE RETURNED ON (B)(4). FCN LEVEL: 1416, SOFTWARE LEVEL: 2.24
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 430316 | AUTOCAT2 WAVE | INTRA-AORTIC BALLOON PUMP | DSP | ARROW INTERNATIONAL INC. | 0001055365 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |