ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2016-14954
- Event Type
- Malfunction
- Date Received
- July 7, 2016
- Report Date
- June 21, 2016
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 09/16/2016. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/23/2016 WITH THE FOLLOWING FINDINGS: THE BLACK BOX SHOWS PUMP WAS RUNNING ON INCORRECT YEAR OF 2007. A MANUAL TIME/DATE CHANGE WAS OBSERVED IN THE BB FROM (B)(6) 2007 19:27 TO (B)(6) 2016 19:29. PUMP WAS EXERCISED FOR 24HRS WITH A 2.0U/HR BASAL RATE; AT END TESTING THE BASAL HISTORY CORRECTLY SHOWED 2.0U AND TDD SHOWED 48.0U. NO DELIVERY INTERRUPTIONS OR EAW¿S OCCURRED DURING THE INVESTIGATION. THE DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECT USER'S PROGRAMMED BASAL RATES. PUMP PASSED DELIVERY ACCURACY TEST AND WAS FOUND TO BE DELIVERING ACCURATELY AND WITHIN REQUIRED RANGE. INVESTIGATORS WERE UNABLE TO DUPLICATE THE COMPLAINT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS ISSUE. THE REPORTER STATED THAT THE BASAL HISTORY DOES NOT MATCH ACTIVE BASAL PROGRAM. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 431416 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |