MYAIRVO HUMIDIFIER
Report
- Report Number
- 9611451-2016-00312
- Event Type
- Malfunction
- Date Received
- July 6, 2016
- Report Date
- June 14, 2016
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K131895
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT MYAIRVO HUMIDIFIER WAS RECEIVED AT OUR FISHER & PAYKEL HEALTHCARE (FPH) REGIONAL OFFICE IN (B)(4) WHERE IT WAS INSPECTED BY A TRAINED FPH TECHNICIAN. THE DEVICE WAS PERFORMANCE TESTED AND THE AUDIBLE ALARM FUNCTION WAS CHECKED. RESULTS: DURING TESTING IT WAS FOUND THAT THE VISUAL ALERTS ON THE COMPLAINT AIRVO WERE OPERATING, HOWEVER NO AUDIBLE ALARM WAS HEARD. PREVIOUS INVESTIGATIONS INTO THIS TYPE OF FAILURE HAVE IDENTIFIED THAT THE PROBLEM IS CAUSED BY A FAULTY SPEAKER AND ELECTRICAL RESISTANCE TESTING HAS SHOWN THE SPEAKER'S RESISTANCE TO BE OPEN CIRCUIT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 150227. CONCLUSION: THE SUPPLIER OF THE SPEAKER UNIT WAS NOTIFIED AND THEY HAVE CARRIED OUT AN INVESTIGATION. THE PROBLEM HAS BEEN TRACED TO AN ISSUE WITH THE GLUING PROCESS. THE SUPPLIER HAS TAKEN STEPS TO ENSURE THAT EACH SPEAKER IS CHECKED FOLLOWING THE GLUING PROCESS AND ANY FOUND FAULTY ARE DISCARDED. AS PART OF OUR ONGOING PRODUCT IMPROVEMENT INITIATIVES, WE RECENTLY IMPLEMENTED A SOAK TEST FOR 100% TESTING OF THE SPEAKER HARNESS ON THE AIRVO PRODUCTION LINE, WHICH IDENTIFIES AND DISCARDS ANY FAULTY SPEAKERS PRIOR TO ASSEMBLY INTO THE AIRVO. THE SUBJECT AIRVO WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THE SOAK TEST. THE MYAIRVO USER MANUAL STATES THAT THE "AIRVO IS FOR THE TREATMENT OF SPONTANEOUSLY BREATHING PATIENTS WHO WOULD BENEFIT FROM RECEIVING HIGH-FLOW, WARMED AND HUMIDIFIED RESPIRATORY GASES." AND THAT " THE UNIT IS NOT INTENDED FOR LIFE SUPPORT." THE USER MANUAL WARNS THE USER: PRIOR TO EACH PATIENT USE, ENSURE THAT THE AUDITORY ALARM SIGNAL IS AUDIBLE BY CONDUCTING THE ALARM SYSTEM FUNCTIONALITY CHECK DESCRIBED IN THE ALARMS SECTION THE ALARM SYSTEM FUNCTIONALITY CHECK INSTRUCTS THE USER ON HOW TO TEST THE ALARM AND STATES THAT "IF EITHER ALARM SIGNAL IS ABSENT, DO NOT USE THE UNIT. CONTACT YOUR FISHER & PAYKEL HEALTHCARE REPRESENTATIVE."
A DISTRIBUTOR IN (B)(6) REPORTED THAT A PT100 MYAIRVO HUMIDIFIER DID NOT HAVE AN AUDIBLE ALARM. THIS WAS FOUND BEFORE PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 427318 | MYAIRVO HUMIDIFIER | MYAIRVO HUMIDIFIER | BTT | FISHER & PAYKEL HEALTHCARE LTD | PT100 | 150227 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |