FDA Adverse Event Malfunction Summary report: N

ENVISION FLEX+, MOUSE, HIGH PH (LINK)

MDR report key: 5773482 · Received July 6, 2016

Report

Report Number
9610099-2016-00007
Event Type
Malfunction
Date Received
July 6, 2016
Date of Event
June 14, 2016
Report Date
July 6, 2016
Manufacturer
DAKO DENMARK A/S
Product Code
LDT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

BASED ON CUSTOMER COMPLAINTS, IT WAS DISCOVERED THAT THE PH OF THE TARGET RETRIEVAL SOLUTION MEASURED 4-5 INSTEAD OF THE CORRECT PH OF 9.1. THE CUSTOMER COMPLAINT WAS RELATED TO NO STAINING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
427768 ENVISION FLEX+, MOUSE, HIGH PH (LINK) ENVISION FLEX+, MOUSE, HIGH PH (LINK) LDT DAKO DENMARK A/S K8002

Patients

Seq Age Sex Outcome Treatment
1 Other