AMISTEM H, HA COATED STEM SIZE 4 STD
Report
- Report Number
- 3005180920-2016-00327
- Event Type
- Injury
- Date Received
- July 6, 2016
- Date of Event
- June 6, 2016
- Report Date
- July 6, 2016
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- PMA / PMN Number
- K093944
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER
Narratives
ON 24 JUNE 2016 THE MEDICAL AFFAIRS DIRECTOR PERFORMED A CLINICAL EVALUATION AND COMMENTED AS FOLLOWS: ASEPTIC LOOSENING OF CEMENTLESS THA, THE TIME IN SITU IS NOT KNOWN, BUT FROM THE ONE RADIOGRAPH AVAILABLE IT MAY BE SURMISED THAT SOME YEARS HAVE PASSED. CORTICAL HYPERTROPHY (MODERATE) IN GRUEN ZONE 4, AND CORTICAL THINNING IN GRUEN ZONES 6 AND 7. THE STEM MAY HAVE SLIGHTLY SUBSIDIZED OR THE INVOLVED LIMB WAS A LITTLE SHORTER. ASEPTIC LOOSENING IS A POSSIBLE KNOWN COMPLICATION FOR CEMENTLESS THA, NO PARTICULAR SIGN OF A DEFECTIVE IMPLANT IS DISCERNIBLE WITH THE INFORMATION AVAILABLE. ON 27 JUNE 2016 THE R&D PROJECT MANAGER ANALYSED THE RETURNED IMPLANTS: OBSERVING THE FEMORAL STEM NO PARTICULAR SIGN CAN BE NOTED, EXCEPT ON THE NECK: SUCH SIGNS WERE PROBABLY CAUSED DURING THE REMOVAL PHASE. THE HA ON THE SURFACE OF THE STEM IS STILL PRESENT IN THE PROXIMAL AREA. ON THE FEMORAL HEAD NO PARTICULAR SIGNS CAN BE SEEN, EXCEPT SOME SCRATCHES PROBABLY DUE TO THE REMOVAL PHASE. IT IS NOT POSSIBLE FROM THE INSPECTION OF THE IMPLANTS DETERMINE THE ROOT CAUSE OF THE EVENT. BATCH REVIEW PERFORMED ON 05 JULY 2016. LOT 103440: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 17 NOVEMBER 2010. EXPIRATION DATE: 2015-10-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.
THE SURGEON HAD TO REVISE THE STEM DUE TO LOOSENING. X-RAYS AND EXPLANTS ARE AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 425454 | AMISTEM H, HA COATED STEM SIZE 4 STD | CEMENTLESS FEMORAL STEM | LZO | MEDACTA INTERNATIONAL SA | 103440 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention |