FDA Adverse Event Malfunction Summary report: N

VENTED HUMIDIFICATION CHAMBER

MDR report key: 5770703 · Received July 5, 2016

Report

Report Number
9611451-2016-00307
Event Type
Malfunction
Date Received
July 5, 2016
Date of Event
June 3, 2016
Report Date
June 10, 2016
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED A HORIZONTAL CRACK AROUND THE BASE OF THE CHAMBER DOME, BETWEEN THE HINGE BRACKET AND ONE OF THE PORTS. SMEARED PRINT WAS ALSO OBSERVED ABOVE THE CRACK LOCATION. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR LOT 150923. CONCLUSION: THE NATURE OF THE CRACKING AND THE SMEARED PRINT INDICATES THAT THE CHAMBER CAME INTO CONTACT WITH A SOLUTION CONTAINING ETHANOL/ALCOHOL, WHICH HAS RESULTED IN ENVIRONMENTAL STRESS CRACKING OF THE CHAMBER DOME. EVERY MR290 CHAMBER IS PRESSURE TESTED TO 200 CMH2O FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DOME DUE TO CRACKS AND OTHER CAUSES. ANY CHAMBER WHICH FAILS THIS TEST IS REJECTED. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER STATE THE FOLLOWING: DO NOT SOAK, WASH, STERILIZE, OR REUSE THIS PRODUCT. AVOID CONTACT WITH CHEMICALS, CLEANING AGENTS, OR HAND SANITIZERS; SET APPROPRIATE VENTILATOR ALARMS; PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; MAXIMUM OPERATING PRESSURE: 8 KPA.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT A CRACK WAS FOUND ON AN MR290V HUMIDIFICATION CHAMBER AFTER 5 DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
422764 VENTED HUMIDIFICATION CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 1509230932

Patients

Seq Age Sex Outcome Treatment
1