AIRVO2 HUMIDIFIER
Report
- Report Number
- 9611451-2016-00299
- Event Type
- Malfunction
- Date Received
- July 4, 2016
- Report Date
- June 6, 2016
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K131895
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT AIRVO HUMIDIFIER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED AND ELECTRICALLY TESTED. RESULTS: DURING TESTING THE AIRVO TURNED ON AND FUNCTIONED, HOWEVER NO AUDIBLE ALARM WAS HEARD. THE FAULT WAS TRACED TO A FAULTY SPEAKER AND ELECTRICAL RESISTANCE TESTING SHOWED THAT THE SPEAKER'S RESISTANCE WAS OPEN CIRCUIT. CONCLUSION: THE SUPPLIER OF THE SPEAKER UNIT WAS NOTIFIED AND THEY HAVE CARRIED OUT AN INVESTIGATION. THE PROBLEM HAS BEEN TRACED TO AN ISSUE WITH THE GLUING PROCESS. THE SUPPLIER HAS TAKEN STEPS TO ENSURE THAT EACH SPEAKER IS CHECKED FOLLOWING THE GLUING PROCESS AND ANY FOUND FAULTY ARE DISCARDED. AS PART OF OUR ONGING PRODUCT IMPROVEMENT INITIATIVES, WE RECENTLY IMPLEMENTED A SOAK TEST FOR 100% TESTING OF THE SPEAKER HARNESS ON THE AIRVO PRODUCTION LINE, WHICH IDENTIFIES AND DISCARDS ANY FAULTY SPEAKERS PRIOR TO ASSEMBLY INTO THE AIRVO. THE SUBJECT AIRVO WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THE SOAK TEST. THE AIRVO USER MANUAL STATES THAT THE "AIRVO IS FOR THE TREATMENT OF SPONTANEOUSLY BREATHING PATIENTS WHO WOULD BENEFIT FROM RECEIVING HIGH-FLOW, WARMED AND HUMIDIFIED RESPIRATORY GASES." AND THAT " THE UNIT IS NOT INTENDED FOR LIFE SUPPORT." THE USER MANUAL WARNS THE USER: PRIOR TO EACH PATIENT USE, ENSURE THAT THE AUDITORY ALARM SIGNAL IS AUDIBLE BY CONDUCTING THE ALARM SYSTEM FUNCTIONALITY CHECK DESCRIBED IN THE ALARMS SECTION THE ALARM SYSTEM FUNCTIONALITY CHECK INSTRUCTS THE USER HOW TO CHECK THE ALARM AND STATES THAT "IF EITHER ALARM SIGNAL IS ABSENT, DO NOT USE THE UNIT. CONTACT YOUR FISHER & PAYKEL HEALTHCARE REPRESENTATIVE."
A HOSPITAL IN (B)(6) REPORTED THAT THE AUDIBLE ALARM OF AN AIRVO 2 HUMIDIFIER WAS NOT FUNCTIONING. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 421932 | AIRVO2 HUMIDIFIER | AIRVO2 HUMIDIFIER | BTT | FISHER & PAYKEL HEALTHCARE LTD | PT101 | 150818 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |