FDA Adverse Event
Other
Summary report: N
KODAK DIRECTVIEW CR CASSETTE
MDR report key: 576710
·
Received February 28, 2005
Report
- Report Number
- 1317307-2005-00001
- Event Type
- Other
- Date Received
- February 28, 2005
- Date of Event
- January 4, 2005
- Report Date
- February 25, 2005
- Manufacturer
- EASTMAN KODAK COMPANY
- Product Code
- MQB
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
A COMPUTED RADIOGRAPHY (CR) SCREEN ARTIFACT RESEMBLING A 3/16" CIRCULAR WHITE DOT WAS NOTED ON 3 CLINICAL IMAGES USING THE KODAK DIRECTVIEW CR CASSETTE /35 X 43 CM WITH ATTACHED STORAGE PHOSPHOR (SP) SCREEN. THE WHITE DOT ARTIFACT WAS IDENTIFIED BY THE RADIOLOGIST AS AN IMAGE ARTIFACT AS IT WAS VISIBLE IN THE SAME LOCATION ON EACH OF THE 3 CLINICIAL IMAGES. AN INVESTIGATION WAS INITIATED ON THIS CUSTOMER COMPLAINT BECAUSE THE RADIOLOGIST HAD CONCERN THAT THIS WHITE DOT ARTIFACT COULD HAVE BEEN MISTAKEN FOR A LUNG NODULE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KODAK DIRECTVIEW CR CASSETTE | CR CASSETTE WITH ATTACHED SP SCREEN | MQB | EASTMAN KODAK COMPANY | NA | 1706 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR | Other |