FDA Adverse Event Other Summary report: N

KODAK DIRECTVIEW CR FLEXIBLE PHOSPHOR SCREEN

MDR report key: 576668 · Received February 28, 2005

Report

Report Number
1317307-2005-00003
Event Type
Other
Date Received
February 28, 2005
Date of Event
February 3, 2005
Report Date
February 25, 2005
Manufacturer
EASTMAN KODAK COMPANY
Product Code
MQB
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

A COMPUTED RADIOGRAPHY (CR) 500 SCREEN ARTIFACT WAS REPORTED BY THE CUSTOEMR SITE TO RESEMBLE A HYPERTROPHIC CALCIFICATION IN THE HIP AREA (TUMOR). THE RADIOLOGIST WAS ABLE TO DETERMINE THAT THIS WAS AN ARTIFACT AS IT APPEARED IN OTHER IMAGES IN THE SAME LOCATION. THERE WERE NO REPORTS OF PT INJURY. KODAK HAS REVIEWED THE CLINICAL AND FLAT FIELD IMAGES FROM THIS CUSTOMER SITE. THE TYPE OF ARTIFACT RESEMBLES AN IRREGLAR SHAPE. IT WAS ALSO REPORTED THAT AN IMAGE ARTIFACT RESEMBLED A "RARE TUMOR" ON A CHEST IMAGE, AND UPON REVIEW OF OTHER IMAGES THE RADIOLOGIST NOTED THE SAME ARTIFACT (RESEMBLING THE "RARE TUMOR") IN THE SAME LOCATIONS AS THE FIRST IMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KODAK DIRECTVIEW CR FLEXIBLE PHOSPHOR SCREEN CR STORAGE PHOSPHOR SCREEN MQB EASTMAN KODAK COMPANY NA 1805

Patients

Seq Age Sex Outcome Treatment
1 YR Other