NAIL HOLDING SCREW, TIBIA T2 TIBIA 10 MM
Report
- Report Number
- 0009610622-2016-00314
- Event Type
- Malfunction
- Date Received
- June 27, 2016
- Date of Event
- June 13, 2016
- Report Date
- June 13, 2016
- Manufacturer
- STRYKER TRAUMA KIEL
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- NURSE
Narratives
THE EVALUATION REVEALED BOTH NAIL HOLDING SCREWS (NHS) AND BOTH NAIL HANDLES TO BE THE PRIMARY PRODUCTS. NO DEVIATIONS WERE FOUND DURING REVIEW OF THE MANUFACTURING AND INSPECTION DOCUMENTS (DHR). THE ITEMS RETURNED WERE DOCUMENTED AS FAULTLESS PRIOR TO DISTRIBUTION. DIMENSIONAL EXAMINATION REVEALED NO DEVIATIONS IN THE RELEVANT UNDAMAGED AREAS. A HARDNESS TEST ACCORDING TO ROCKWELL REVEALED THE HARDNESS OF THE MATERIAL OF ALL ITEMS BEING WITHIN SPECIFIED PARAMETERS. BOTH NHS STUCK WITHIN THE NAIL HANDLES, SO THAT THE NAIL WAS NOT DETACHABLE FROM THE HANDLE WITH CODE #3641. IT WAS NECESSARY TO CUT THE NAIL OFF AND REMOVE THE NHS USING A HAMMER AND A MALLET. AFTER DISASSEMBLING SIGNIFICANT FRETTING MARKS WERE VISIBLE, RUNNING OVER THE WHOLE CIRCUMFERENCE ON THE OUTER SURFACE OF THE NHS SHAFTS, CLOSE TO THE BEGINNING OF THE THREAD. CORRESPONDING FRETTING MARKS WERE ALSO VISIBLE, RUNNING OVER THE WHOLE CIRCUMFERENCE INSIDE THE TUBE OF THE NAIL HANDLES. THE FRETTING CAUSED THE JAMMING OF THE INSTRUMENTS. FRETTING IS A RARE BUT KNOWN REACTION IN CASE THESE MATERIALS COME IN CONTACT WHICH EACH OTHER. DURING SCREWING INTO THE NAIL IT IS POSSIBLE THAT THE NHS GETS IN CONTACT WITH THE INNER SURFACE OF THE NAIL HANDLE TUBE AND FRICTION OCCURS DUE TO A SUBOPTIMAL (SLIGHT OBLIQUE) AXIAL INSERTION. THE APPEARANCE OF THE DAMAGE INDICATED THAT THE DEVICES GOT JAMMED BY ¿COLD WELDING¿ DUE TO FRICTION CORROSION BEING CAUSED BY HIGH SURFACE PRESSURE. MOST LIKELY HIGH SURFACE PRESSURE GENERATED BY HIGH LOAD APPLIED TO THE NHS HAD LED TO THE FOUND DAMAGES. LOCAL SURFACE PRESSURE MAY ARISE WHEN THE ITEMS ARE ASSEMBLED UNDER MISALIGNMENT (USE ERROR). A PROCESS CHANGE WAS ALREADY PERFORMED IN 2004 TO PREVENT FRETTING REGARDING THE NAIL HOLDING SCREW (SURFACE COATING WAS REMOVED). NO FURTHER ACTIONS WERE INITIATED. THE RETURNED NHS WERE MANUFACTURED POST TO THE CHANGE. THE CHANGE DOES NOT PREVENT A USER ERROR DUE TO OBLIQUE INSERTION. ADDITIONALLY HITTING MARKS WERE FOUND ON BOTH NAIL HANDLE SURFACES. HAMMERING THE NAIL HANDLES IS NOT ALLOWED (ABNORMAL USE). THE HANDLES HAVE THREADED SLOTS TO ASSEMBLE A STRIKE PLATE. REVIEW OF COMPLAINT HISTORY, CAPA DATABASES AND RISK ANALYSIS DID NOT IDENTIFY ANY CONSPICUITY. THE REVIEW OF THE RISK ASSESSMENT FOR THE FAILURE MODE INDICATED THE ISSUE WAS ADDRESSED ADEQUATELY. NO NON-CONFORMITY WAS IDENTIFIED.
ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
THE CUSTOMER REPORTED THAT THE HOLDING SCREW T2 TIBIA GOT STUCK IN NAIL ADAPTER TWICE. THE CUSTOMER REPORTED THAT THE SECOND NAIL COULD BE INSERTED FREE-HAND. NO PATIENT HARM WAS DONE. THE STRYKER SALES REP HAS CONFIRMED THAT THERE ARE 2 INSTRUMENTS WITH THE SAME REPORTED ISSUE IN THE ONE OPERATION.
THE CUSTOMER REPORTED THAT THE HOLDING SCREW T2 TIBIA GOT STUCK IN NAIL ADAPTER TWICE. THE CUSTOMER REPORTED THAT THE SECOND NAIL COULD BE INSERTED FREE-HAND. NO PATIENT HARM WAS DONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 408266 | NAIL HOLDING SCREW, TIBIA T2 TIBIA 10 MM | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | STRYKER TRAUMA KIEL | KSS148120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |