COBE CENTRYSYSTEM 3
Report
- Report Number
- 1713683-1996-00287
- Event Type
- Malfunction
- Date Received
- December 20, 1996
- Report Date
- December 20, 1996
- Manufacturer
- GAMBRO HEALTHCARE
- Product Code
- FII
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
RETURNED FOUR-POSITION LINE CLAMP WAS VISUALLY INSPECTED AND CONFIRMED THAT CLAMP HOLDER WAS BROKEN. TABS ON WHITE CLAMP INSERT HOLDER WAS BROKEN OR STRETCHED, ALLOWING INSERTS TO FALL OUT OF HOLDER. IT WAS ALSO DETERMINED THAT RETURNED CLAMP STOP TABS WHICH HOLD CLAMP IN OCCULUDED POSTION WERE EITHER WORN OR BROKEN, SO THAT CLAMP NO LONGER STOPPED SECURELY IN CLOSED POSITION. THIS CONDITION OCCURRED WITH NORMAL USE. SAFETY ANALYSIS: IF CLAMP FAILS, OPERATOR CAN CLAMP LINE WITH A HEMOSTAT OR INSTALL A LUER CAP ONTO LINE. OPERATOR'S MANUAL, VERSION 1995/10, RECOMMENDS IN SECTION 3-43 THAT LINES GOING THROUGH LINE CLAMP BE DOUBLE CLAMPED TO ELIMINATE ANY UNWANTED LEAKAGE, EVEN WHEN MACHINE CLAMP IS OPERATING PROPERLY. ALSO, IT SHOULD BE OBVIOUS TO USER NOT TO USE CLAMP IF IT IS BROKEN. HOWEVER, IF A FACILITY CHOOSES TO USE CLAMP WHEN BROKEN AND NOT DOUBLE CLAMP, UNANTICIPATED SALINE LEAKS TO PT CAN RESULT. ABOVE INFO INDICATES THAT LINE CLAMP FAILURE DESCRIBED IN THIS COMPLAINT WAS GENERATED BY A BROKEN CLAMP ASSEMBLY. THIS ISSUE WILL CONTINUE TO BE TRENDED AS PART OF CO'S CORRECTIVE ACTION SYSTEM AND MANAGEMENT REVIEW TEAM. ANY FURTHER INVESTIGATIONS, ANALYSES, AND/OR CORRECTIVE ACTIONS TAKEN ON THIS COMPLAINT ISSUE WILL BE DETERMINED BY AND DOCUMENTED IN THIS CORRECTIVE ACTION REVIEW PROCESS. SEE FAR #960027.
DURING A DIALYSIS TREATMENT, A FOUR POSITION LINE CLAMP WAS MISSING TWO OF THE ARMS. THERE WAS NO PT INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBE CENTRYSYSTEM 3 | DIALYSIS CONTROL UNIT | FII | GAMBRO HEALTHCARE | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |