FDA Adverse Event Other Summary report: N

BD LOGIC BLOOD GLUCOSE MONITOR

MDR report key: 575185 · Received February 10, 2005

Report

Report Number
2243072-2005-00037
Event Type
Other
Date Received
February 10, 2005
Date of Event
January 15, 2005
Report Date
February 10, 2005
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT GETTING ERRATIC READINGS WITH THEIR LOGIC METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVERAGE READINGS (326) AS REF. POINT VS. BD READINGS OF 143, 599, 237 YIELDED DATA POINTS IN THE D/C ZONE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD LOGIC BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 4064181

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other