FDA Adverse Event Malfunction Summary report: N

COBE CENTRYSYSTEM 3

MDR report key: 57492 · Received December 20, 1996

Report

Report Number
1713683-1996-00279
Event Type
Malfunction
Date Received
December 20, 1996
Report Date
December 17, 1996
Manufacturer
GAMBRO HEALTHCARE
Product Code
FII
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE WERE NO RELATED CPF DEFECTS REPORTED BY THIS MACHINE. RETURNED FOUR-POSITION LINE CLAMP WAS VISUALLY INSPECTED AND CONFIRMED THAT CLAMP HOLDER WAS BROKEN. TABS ON WHITE CLAMP INSERT HOLDER WAS BROKEN OR STRETCHED, ALLOWING INSERTS TO FALL OUT OF HOLDER. IT WAS ALSO DETERMINED THAT RETURNED CLAMP STOP TABS WHICH HOLD CLAMP IN OCCLUDED POSTION WERE EITHER WORN OR BROKEN, SO THAT CLAMP NO LONGER STOPPED SECURELY IN CLOSED POSTION. THIS CONDITION OCCURRED WITH NORMAL USE. ABOVE INFO INDICATES THAT LINE CLAMP FAILURE DESCRIBED IN THIS COMPLAINT WAS GENERATED BY A BROKEN CLAMP ASSEMBLY. THIS ISSUE WILL BE TRENDED AS PART OF GAMBRO HEALTHCARE CORRECTIVE ACTION SYSTEM AND MANAGEMENT REVIEW TEAM. ANY FURTHER INVESTIGATIONS, ANALYSES, AND/OR CORRECTIVE ACTIONS TAKEN OF THIS COMPLAINT ISSUE WILL BE DETERMINED BY AND DOCUMENTED IN THIS CORRECTIVE ACTION REVIEW PROCESS. SEE FAR# 960027. CUSTOMER CONTACTED: NO. SALES/SERVICE CONTACTED BY INVESTIGATOR? NO. TRAINING REQUIRED? NO.

Description of Event or Problem · 1

DURING A DIALYSIS TREATMENT, THE CLAMPS ON THE FOUR POSITION LINE CLAMP WERE FALLING OFF. THERE WAS NO PT INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBE CENTRYSYSTEM 3 DIALYSIS CONTROL UNIT FII GAMBRO HEALTHCARE NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other