INRATIO PT/INR TEST STRIPS
Report
- Report Number
- 2027969-2016-00469
- Event Type
- Malfunction
- Date Received
- June 23, 2016
- Date of Event
- May 4, 2016
- Report Date
- June 6, 2016
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K110212
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
IT IS INDICATED THAT THE PRODUCT IS NOT RETURNING FOR EVALUATION. THEREFORE, A REVIEW OF IN-HOUSE TESTING WAS PERFORMED. IN-HOUSE STRIP TESTING ON STRIP LOT K380861 MEETS CRITERIA. THE PRODUCT PERFORMED AS EXPECTED. A REVIEW OF THE MANUFACTURING RECORDS FOR THE LOT WAS PERFORMED; THE LOT MET RELEASE SPECIFICATION. UNABLE TO DETERMINE THE ROOT CAUSE OF THE CUSTOMER'S COMPLAINT. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
THE CUSTOMER DID NOT PROVIDE A LOT NUMBER OR RETURN ANY PRODUCTS FOR INVESTIGATION. SINCE THE PRODUCT(S) ASSOCIATED WITH THE COMPLAINT WAS NOT RETURNED AND A LOT NUMBER WAS NOT PROVIDED, MANUFACTURING RECORD REVIEW COULD NOT BE PERFORMED AND FURTHER INVESTIGATION WAS NOT POSSIBLE.
THE LOT NUMBER K380861 WAS NOT INCLUDED IN FOLLOW UP #1 ALTHOUGH THIS INFORMATION WAS PROVIDED PRIOR TO THE FOLLOW UP BEING SUBMITTED. HOWEVER, THE INFORMATION DOES INCLUDE THE INVESTIGATION PERFORMED FOR THIS LOT.
A CALLER REPORTED A VARIANCE BETWEEN THE INRATIO INR RESULT AND ALTERNATE POC INR RESULT. THE RESULTS WERE AS FOLLOWS: (B)(6) 2016: INRATIO INR=2.7, (B)(6) 2016: ALTERNATE POC INR=1.8. THE REPORTED THERAPEUTIC RANGE IS: 2.0-2.5.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 400862 | INRATIO PT/INR TEST STRIPS | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | HS99007G1 | K380861 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |