FDA Adverse Event Malfunction Summary report: N

ADVIA 2120I WITH SINGLE ASPIRATE AUTOSAMPLER

MDR report key: 5743297 · Received June 22, 2016

Report

Report Number
2432235-2016-00334
Event Type
Malfunction
Date Received
June 22, 2016
Date of Event
May 8, 2016
Report Date
June 22, 2016
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
GKZ
PMA / PMN Number
K102644
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER STATED SHE FOLLOWED THE PROCEDURE IN THE ADVIA 2120I OPERATOR GUIDE WHEN PERFORMING THE CENTERING COLLAR MAINTENANCE ON THE ADVIA 2120I WITH SINGLE ASPIRATE AUTOSAMPLER INSTRUMENT. SHE WAS INJURED WHILE REMOVING THE CENTERING COLLAR AND BEFORE THE RED NEEDLE GUARD WAS PUT IN PLACE OVER THE AUTOSAMPLER NEEDLE. SIEMENS HEALTHCARE DIAGNOSTICS INC. HAS DETERMINED THAT THE CAUSE OF THE NEEDLESTICK INJURY TO THE FINGER WHILE PERFORMING THE CENTERING COLLAR MAINTENANCE ON THE ADVIA 2120I WITH SINGLE ASPIRATE AUTOSAMPLER INSTRUMENT IS OPERATOR'S TECHNIQUE. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.

Description of Event or Problem · 1

THE CUSTOMER SUSTAINED A NEEDLE STICK INJURY TO THEIR FINGER WHILE PERFORMING THE CENTERING COLLAR MAINTENANCE ON THE ADVIA 2120I WITH SINGLE ASPIRATE AUTOSAMPLER INSTRUMENT. THE INCIDENT OCCURRED AS THE CUSTOMER WAS REMOVING THE CENTERING COLLAR DURING THE MAINTENANCE PROCEDURE. THE CUSTOMER WAS WEARING A LAB COAT AND GLOVES AT THE TIME OF THE INCIDENT. AS PER THE HOSPITAL'S LAB PROCEDURE, THE CUSTOMER VISITED THE ACCIDENT AND EMERGENCY DEPARTMENT IN THE HOSPITAL AND THE WOUND WAS ENCOURAGED TO BLEED. THERE WAS NO PROPHYLACTIC TREATMENT RECEIVED. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO THE CUSTOMER BEING INJURED WHILE PERFORMING THE CENTERING COLLAR MAINTENANCE ON THE ADVIA 2120I WITH SINGLE ASPIRATE AUTOSAMPLER INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
396103 ADVIA 2120I WITH SINGLE ASPIRATE AUTOSAMPLER ADVIA 2120I WITH SINGLE ASPIRATE AUTOSAMPLER GKZ SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 2120I WITH SINGLE ASPIRATE AUTOSAMPLER

Patients

Seq Age Sex Outcome Treatment
1