OT VERIO2 METER
Report
- Report Number
- 3008382007-2016-30997
- Event Type
- Malfunction
- Date Received
- June 22, 2016
- Report Date
- June 14, 2016
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K131363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PATIENT
Narratives
THE LAY USER/PATIENTS HAVE RETURNED 2 LOTS OF TEST STRIPS (LOT # 3955248 AND 3966711). THESE HAVE BEEN EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: BOTH LOTS OF TEST STRIPS PASSED TESTING; THE REPORTED ISSUE COULD NOT BE CONFIRMED. HOWEVER, SECONDARY AND TERTIARY ISSUES WERE NOTED IN TEST STRIP LOT # 3955248. WHEN TEST STRIPS WERE TESTED WITH CONTROL SOLUTION, AN ERROR 4 WAS OBSERVED WITH NO ASSIGNABLE CAUSE FOUND. THE TEST STRIPS WERE ALSO FOUND TO HAVE RESULTS BELOW RANGE WHEN TESTED WITH CONTROL SOLUTION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2016, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING ERROR 2. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 397977 | OT VERIO2 METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |