PRODIGY AUTOCODE
Report
- Report Number
- 3005862821-2016-00032
- Event Type
- Injury
- Date Received
- June 21, 2016
- Date of Event
- May 12, 2016
- Report Date
- May 20, 2016
- Manufacturer
- OK BIOTECH CO., LTD.
- Product Code
- NBW
- PMA / PMN Number
- K073118
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
SUSPECTED DEVICE EVALUATED BY OK BIOTECH AND CALCULATED THAT THE METER OPERATED WITHIN SPECIFICATIONS. THE STANDBY CURRENT TEST IS 16.2?A. THE CRITERIA IS <55?A. PASS. METER SETTING, AUDIO AND ALL BUTTONS FUNCTION ARE OK. PATIENT SENT BACK THE SUSPECTED STRIP, STRIP LOT# D150603-2. WE TESTED THE RETURNED STRIP RETURNED STRIPS, THE CONTROL SOLUTION TESTS FOR LEVEL LOW WERE 45/46 MG/DL; FOR LEVEL HIGH WERE 252/261 MG/DL. THE REQUEST CONTROL SOLUTION RANGES ARE: LEVEL LOW 25~70 MG/DL; LEVEL HIGH 210~310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS 4. WE TESTED THE SUSPECTED METER WITH IN HOUSE CONTROL SOLUTION AND SAME LOT NUMBER OF THE RETAIN STRIPS THAT PATIENT SENT BACK TO US. THE CONTROL SOLUTION TESTS FOR LEVEL LOW ARE 49/49 MG/DL; FOR LEVEL HIGH ARE 273/274 MG/DL. THE CONTROL SOLUTION RANGES ARE: LEVEL LOW 25~70 MG/DL; LEVEL HIGH 210~310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS.
OUR IMPORTER, (B)(4), RECEIVED A CALL ON 05/12/2016 REPORTING A MEDICAL INTERVENTION THAT OCCURRED ON (B)(6) 2016 AT 6:50AM. PATIENT'S GRANDDAUGHTER CALLED IN STATING THAT PATIENT WAS ACTING STRANGELY AND MAKING STATEMENTS THAT DID NOT MAKE SENSE (LETHARGIC). SHE TESTED PATIENT'S GLUCOSE AND RECEIVED A READING OF 143MG/DL WITH THE PRODIGY METER. PATIENT'S DAUGHTER THEN CALL MEDICS AND UPON ARRIVAL MEDICS TESTED PATIENT'S GLUCOSE WITH A READING OF 40MG/DL. NO MEDICATIONS WERE TAKEN PRIOR TO THE EVENT. PATIENT CONSUMED 1 CUP OF ORANGE JUICE PRIOR TO THE MEDICAL ATTENTION EVENT. PATIENT'S NORMAL GLUCOSE READING AROUND THE TIME OF THE DAY OF THE EVENT IS 90-100MG/DL. PARAMEDICS WERE CALLED 5 MINUTES AFTER THE EVENT. PATIENT WAS GIVEN 1 CUP OF ORANGE JUICE WHILE WAITING ON MEDICS TO ARRIVE. MEDICS ARRIVED 10 MINUTES AFTER BEING CALLED. PARAMEDICS ADMINISTERED GLUCOSE AND ORANGE JUICE TO PATIENT. PATIENT WAS NOT TRANSPORTED TO ER. (B)(4) SENT REPLACEMENT AND PREPAID ENVELOPE REQUESTING RETURN OF SUSPECT SYSTEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 393480 | PRODIGY AUTOCODE | BLOOD GLUCOSE MONITORING DEVICE | NBW | OK BIOTECH CO., LTD. | 51850 | D150603-2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Required Intervention | GOES TO CANCER CENTER EVERY 6 WEEKS |