FDA Adverse Event Injury Summary report: N

PRODIGY AUTOCODE

MDR report key: 5737846 · Received June 21, 2016

Report

Report Number
3005862821-2016-00032
Event Type
Injury
Date Received
June 21, 2016
Date of Event
May 12, 2016
Report Date
May 20, 2016
Manufacturer
OK BIOTECH CO., LTD.
Product Code
NBW
PMA / PMN Number
K073118
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SUSPECTED DEVICE EVALUATED BY OK BIOTECH AND CALCULATED THAT THE METER OPERATED WITHIN SPECIFICATIONS. THE STANDBY CURRENT TEST IS 16.2?A. THE CRITERIA IS <55?A. PASS. METER SETTING, AUDIO AND ALL BUTTONS FUNCTION ARE OK. PATIENT SENT BACK THE SUSPECTED STRIP, STRIP LOT# D150603-2. WE TESTED THE RETURNED STRIP RETURNED STRIPS, THE CONTROL SOLUTION TESTS FOR LEVEL LOW WERE 45/46 MG/DL; FOR LEVEL HIGH WERE 252/261 MG/DL. THE REQUEST CONTROL SOLUTION RANGES ARE: LEVEL LOW 25~70 MG/DL; LEVEL HIGH 210~310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS 4. WE TESTED THE SUSPECTED METER WITH IN HOUSE CONTROL SOLUTION AND SAME LOT NUMBER OF THE RETAIN STRIPS THAT PATIENT SENT BACK TO US. THE CONTROL SOLUTION TESTS FOR LEVEL LOW ARE 49/49 MG/DL; FOR LEVEL HIGH ARE 273/274 MG/DL. THE CONTROL SOLUTION RANGES ARE: LEVEL LOW 25~70 MG/DL; LEVEL HIGH 210~310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS.

Description of Event or Problem · 1

OUR IMPORTER, (B)(4), RECEIVED A CALL ON 05/12/2016 REPORTING A MEDICAL INTERVENTION THAT OCCURRED ON (B)(6) 2016 AT 6:50AM. PATIENT'S GRANDDAUGHTER CALLED IN STATING THAT PATIENT WAS ACTING STRANGELY AND MAKING STATEMENTS THAT DID NOT MAKE SENSE (LETHARGIC). SHE TESTED PATIENT'S GLUCOSE AND RECEIVED A READING OF 143MG/DL WITH THE PRODIGY METER. PATIENT'S DAUGHTER THEN CALL MEDICS AND UPON ARRIVAL MEDICS TESTED PATIENT'S GLUCOSE WITH A READING OF 40MG/DL. NO MEDICATIONS WERE TAKEN PRIOR TO THE EVENT. PATIENT CONSUMED 1 CUP OF ORANGE JUICE PRIOR TO THE MEDICAL ATTENTION EVENT. PATIENT'S NORMAL GLUCOSE READING AROUND THE TIME OF THE DAY OF THE EVENT IS 90-100MG/DL. PARAMEDICS WERE CALLED 5 MINUTES AFTER THE EVENT. PATIENT WAS GIVEN 1 CUP OF ORANGE JUICE WHILE WAITING ON MEDICS TO ARRIVE. MEDICS ARRIVED 10 MINUTES AFTER BEING CALLED. PARAMEDICS ADMINISTERED GLUCOSE AND ORANGE JUICE TO PATIENT. PATIENT WAS NOT TRANSPORTED TO ER. (B)(4) SENT REPLACEMENT AND PREPAID ENVELOPE REQUESTING RETURN OF SUSPECT SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
393480 PRODIGY AUTOCODE BLOOD GLUCOSE MONITORING DEVICE NBW OK BIOTECH CO., LTD. 51850 D150603-2

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention GOES TO CANCER CENTER EVERY 6 WEEKS