M2A-MAGNUM MOD HD SZ 42MM
Report
- Report Number
- 0001825034-2016-02138
- Event Type
- Injury
- Date Received
- June 20, 2016
- Date of Event
- August 18, 2014
- Report Date
- November 16, 2016
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- KWA
- PMA / PMN Number
- PK042037
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- ATTORNEY
Narratives
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, "MATERIAL SENSITIVITY REACTIONS." NUMBER 6 STATES, "INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS." NUMBER 14 STATES, "INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN." NUMBER 15 STATES, "ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL ON METAL ARTICULATING SURFACES." (B)(4). THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. PRODUCT LOCATION UNKNOWN.
THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.
THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. CONCOMITANT PRODUCTS: BIOMET M2A MAGNUM CUP CATALOG#: US157848 LOT#: 092790, BIOMET TAPERLOC STEM CATALOG#: 103201 LOT#: 340660. (B)(4). THIS REPORT IS NUMBER 1 OF 2 MDRS FILED FOR THE SAME PATIENT (REFERENCE 1825034-2016-02138 & 04717).
LEGAL COUNSEL FOR PATIENT REPORTED THAT PATIENT UNDERWENT A RIGHT HIP REVISION PROCEDURE APPROXIMATELY SEVEN YEARS POST-IMPLANTATION DUE TO ALLEGED PAIN, DISCOMFORT, SORENESS, DYSFUNCTION, LOSS OF RANGE OF MOTION, POPPING SOUND, METAL DEBRIS, AND METALLOSIS. REVIEW OF INVOICE HISTORY CONFIRMS THE MODULAR HEAD WAS REMOVED AND REPLACED WITH A CERAMIC HEAD AND A POLY BEARING WAS IMPLANTED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.
LEGAL COUNSEL FOR PATIENT REPORTED THAT PATIENT UNDERWENT A RIGHT HIP REVISION PROCEDURE APPROXIMATELY SEVEN YEARS POST-IMPLANTATION DUE TO ALLEGED PAIN, DISCOMFORT, SORENESS, DYSFUNCTION, LOSS OF RANGE OF MOTION, POPPING SOUND, METAL DEBRIS, AND METALLOSIS. REVIEW OF INVOICE HISTORY CONFIRMS THE MODULAR HEAD WAS REMOVED AND REPLACED WITH A CERAMIC HEAD AND A POLY BEARING WAS IMPLANTED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. IT WAS REPORTED IN OPERATIVE REPORT RECEIVED THAT PATIENT UNDERWENT A RIGHT HIP REVISION PROCEDURE APPROXIMATELY SEVEN YEARS POST-IMPLANTATION DUE TO PAIN AND SWELLING. DURING THE PROCEDURE, PSEUDOTUMOR FORMATION AND METALLOSIS JOINT FLUID WERE NOTED. THE FEMORAL HEAD WAS REMOVED AND REPLACED. AN ACETABULAR BEARING WAS ALSO IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 391085 | M2A-MAGNUM MOD HD SZ 42MM | PROSTHESIS, HIP | KWA | BIOMET ORTHOPEDICS | N/A | 069140 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Hospitalization| R |