PACS HW SERVER
Report
- Report Number
- 3004526608-2016-00001
- Event Type
- Death
- Date Received
- June 20, 2016
- Date of Event
- May 21, 2016
- Report Date
- July 27, 2016
- Manufacturer
- GE HEALTHCARE IT
- Product Code
- LLZ
- PMA / PMN Number
- K043415
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
PATIENT INFORMATION COULD NOT BE OBTAINED DUE TO COUNTRY PRIVACY LAWS. THE INITIAL REPORTER IS LOCATED OUTSIDE THE U.S., AND THEREFORE THIS INFORMATION IS NOT PROVIDED DUE TO COUNTRY PRIVACY LAWS. MANUFACTURE DATE IS UNKNOWN. GE HEALTHCARE'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
THE INVESTIGATION REVEALED THAT THE MERGING OF TWO PATIENT STUDIES WAS PERFORMED CORRECTLY AND THAT THE STUDY HAD THE APPROPRIATE TIME AND DATE STAMP AFTER THE MERGE. THE INVESTIGATION ALSO REVEALED THAT THE EXAM WAS PRESENT ON THE WORKLIST AND AVAILABLE FOR USE BY THE RADIOLOGIST. THE INVESTIGATION DID NOT IDENTIFY ANY DEFICIENCY WITH THE CPACS PRODUCT. THE CPACS PRODUCT IS WORKING AS DESIGNED.
THE CUSTOMER REPORTED THE PATIENT WAS HOSPITALIZED FOR A SMALL BOWEL OBSTRUCTION AND PAIN. A CT WAS ORDERED FOR THE PATIENT. THE CT SCAN WAS PERFORMED ON (B)(6) 2016. THIS STUDY WAS VIEWED BY RADIOLOGIST ON (B)(6) 2016 WHEN AN EXAM NOTE WAS PERFORMED BY THE RADIOLOGIST THAT INDICATED: SBO WITH FREE AIR AND FLUID. THE WARD DOCTOR WAS NOTIFIED OF THESE FINDINGS ON (B)(6) 2016. THE WARD DOCTOR DETERMINED THAT THE PATIENT WAS TOO FRAIL FOR SURGERY, SO NO SURGERY WAS PERFORMED. THE PATIENT DIED ON (B)(6) 2016. THE FINAL REPORT (SIGNED (B)(6) 2016) INDICATED THAT THERE WAS A DELAY IN READING BECAUSE THE STUDY WAS INCORRECTLY DATE AND TIME STAMPED
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 390262 | PACS HW SERVER | SYSTEM, IMAGE PROCESSING, RADIOLOGIC | LLZ | GE HEALTHCARE IT | IAS001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death |