OT VERIO2 METER
Report
- Report Number
- 3008382007-2016-30223
- Event Type
- Malfunction
- Date Received
- June 18, 2016
- Report Date
- June 10, 2016
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K131363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PATIENT
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2016, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING THAT THE SUBJECT METER READ INACCURATELY HIGH COMPARED TO ANOTHER, UNKNOWN, DEVICE. THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF "155 MG/DL" WITH THE SUBJECT METER BUT WAS UNABLE TO RECALL THE RESULT FROM THE OTHER DEVICE. THE TESTS WERE PERFORMED WITHIN 30 MINUTES OF EACH OTHER. THE REPORTER ALSO CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF "146, 158, 159, 171, 154, 151, 167 AND 155 MG/DL" WITH THE SUBJECT METER, THEY WERE UNABLE TO RECALL THE RESULTS FROM THE OTHER DEVICE. THE TESTS WERE PERFORMED WITHIN AN UNKNOWN TIME OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN'S CRITERIA FOR ACCURACY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.