AEX GENERATOR
Report
- Report Number
- 1226420-2016-00081
- Event Type
- Malfunction
- Date Received
- June 17, 2016
- Date of Event
- May 12, 2016
- Report Date
- August 16, 2016
- Manufacturer
- MEDTRONIC ADVANCED ENERGY
- Product Code
- GEI
- PMA / PMN Number
- K143175
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4): BRIEF DESCRIPTION OF COMPLAINT: THE SURGEON REPORTED A SPARK FROM THE DEVICE, CAUSING THE ICD LEAD TO STAIN. INVESTIGATION PLAN: VISUAL INSPECTION FUNCTIONAL INSPECTION (IF APPLICABLE) SLHR REVIEW COMPLAINT DEVICE DETAILS: PRODUCT CODE (CFN): 40-405-1 SERIAL: (B)(4) TESTING PERFORMED: VISUAL INSPECTION: PACKAGING: GENERATOR WAS RETURNED IN AN APPROVED MAE BOX WITH PROPER FOAM PROTECTION. THE PACKAGING DID NOT APPEAR TO HAVE SUSTAINED DAMAGE DURING TRANSIT. EXTERNAL VISUAL: THERE WAS NO EVIDENCE OF EXTERNAL DEFECTS TO THE GENERATOR. INTERNAL VISUAL: THERE WAS NO EVIDENCE OF INTERNAL DEFECTS TO THE GENERATOR. UNIT WAS NOT OPENED. FUNCTIONAL INSPECTION: TEST(S) PERFORMED IN ACCORDANCE WITH REPORTED ISSUE: FUNCTIONAL TEST, MONOPOLAR POWER OUTPUT TEST (TABLE 1) MEASUREMENTS ARE WITHIN SPECIFICATION. EQUIPMENT USED FOR TESTING: PS210-030S, PS200-040, 23-112-1, 7254, 7255, 7282 RESULTS OF EACH TEST PERFORMED: THE GENERATOR SUCCESSFULLY PASSED ALL REQUIREMENTS FOR BOTH FUNCTIONAL TEST AND MONOPOLAR POWER OUTPUT TEST. NO SPARK WAS OBSERVED DURING TESTING. THE FOLLOWING IMPACTS TO FUNCTIONALITY WERE IDENTIFIED: N/A DEFECTS WERE RESOLVED BY: N/A THERE WERE NO ADDITIONAL FAILURES FOUND DURING ANALYSIS. SLHR REVIEW: THIS GENERATOR HAS NO HISTORY OF SERVICE RELATED TO THE CURRENT FINDINGS. INVESTIGATION CONCLUSION: THE REPORTED ISSUE WAS NOT CONFIRMED. REFERENCE DOCUMENTS: COMPLAINT ANALYSIS-GENERATORS-WORK INSTRUCTIONS, 42-10-1119 REV. B AEX SERVICE PROCESS INSTRUCTIONS , 61-90-0703 REV. E.
PRODUCT EVENT: (B)(4). PATIENT INFORMATION INCOMPLETE AND MISSING PATIENT INFORMATION UNABLE TO BE OBTAINED DESPITE A GOOD FAITH EFFORT MADE TO OBATIN THE INFORMAITON FROM THE CUSTOMER.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
DURING AN ICD EXCHANGE THE SURGEON ACTIVATED THE PLASMABLADE DEVICE AND REPORTED A SPARK ON THE TISSUE CAUSING A STAIN ON ONE OF THE ICD LEADS. THE SURGEON WIPED OFF THE STAIN AND REPORTED NO TISSUE DAMAGE OR IMPACT TO THE LEAD.
DURING AN ICD EXCHANGE THE SURGEON ACTIVATED THE PLASMABLADE DEVICE AND REPORTED A SPARK ON THE TISSUE CAUSING A STAIN ON ONE OF THE ICD LEADS. THE SURGEON WIPED OFF THE STAIN AND REPORTED NO TISSUE DAMAGE OR IMPACT TO THE LEAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 386710 | AEX GENERATOR | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | MEDTRONIC ADVANCED ENERGY | 40-405-1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |