FDA Adverse Event Injury Summary report: N

AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE

MDR report key: 5722405 · Received June 14, 2016

Report

Report Number
2183959-2016-00129
Event Type
Injury
Date Received
June 14, 2016
Date of Event
May 9, 2016
Report Date
May 25, 2016
Manufacturer
AMERICAN MEDICAL SYSTEMS (MN)
Product Code
JCW
PMA / PMN Number
N970012
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED FOR EVALUATION; ANALYSIS RESULTS INDICATE ONE CYLINDER AND THE RESERVOIR PERFORMED WITHIN SPECIFICATIONS. THE OTHER CYLINDER SUSTAINED A HOLE IN THE OUTER TUBE THAT APPEARS TO BE CAUSED BY A SHARP INSTRUMENT; THE CYLINDER REMAINED FUNCTIONAL. THE PUMP WAS NOT FUNCTIONALLY TESTED DUE CONTAMINATION BY WHAT APPEARS TO BE TISSUE.

Additional Manufacturer Narrative · 1

COMMON DEVICE NAME FROM PENILE PROSTHESIS TO DEVICE, IMPOTENCE, MECHANICAL/HYDRAULIC. PRODUCT CODE (FDA) FROM FAE TO FHW. PMA/510(K) FROM K821628 TO N970012 .

Description of Event or Problem · 1

IT WAS REPORTED THAT FOLLOWING THE IMPLANTATION OF HIS AMS 700 INFLATABLE PENILE PROSTHESIS, THE PATIENT EXPERIENCED INFECTION. "ONE MONTH POST-OP ((B)(6)), PATIENT PRESENTED WITH SCROTAL CELLULITIS AND RECEIVED 2 WEEKS OF AUGMENTIN. PATIENT SEEN IN SURGEON'S OFFICE FOR SCROTAL INCISION FLUCTUANCE. PATIENT RETURNED 2 DAYS LATER WITH OPEN INCISION AND DRAINED PURULENT FLUID, SCROTAL REDNESS, FIRMNESS, SCROTUM AND SUPRAPUBIC AREA PAIN WITH SYMPTOMATIC FEVER AND CHILLS. WBC COUNT 12,000. PATIENT WAS HOSPITALIZED AND RECEIVED BROAD-SPECTRUM IV ANTIBIOTICS FOR 72 HOURS. WHITE COUNT NORMALIZED AND CELLULITIS RESOLVED. PATIENT TAKEN TO SURGERY (B)(6) 2016 FOR SALVAGE, WASHOUT WITH REMOVAL OF PROSTHESIS AND WASHOUT AND PLACEMENT OF AN AMS MALLEABLE PROSTHESIS". PATIENT DID WELL POST-OP AND DISCHARGED TO HOME IN STABLE CONDITION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED IN RELATION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379291 AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE DEVICE, IMPOTENCE, MECHANICAL/HYDRAULIC JCW AMERICAN MEDICAL SYSTEMS (MN)

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| R