HAKIM PROGRAMMABLE VALVE MICRO W/RICKHAM
Report
- Report Number
- 1226348-2016-10424
- Event Type
- Malfunction
- Date Received
- June 13, 2016
- Date of Event
- May 19, 2016
- Manufacturer
- CODMAN & SHURTLEFF, INC. / MEDOS S.A.
- Product Code
- JXG
- PMA / PMN Number
- PK980778
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
(B)(4). UPON COMPLETION OF THE INVESTIGATION, IT WAS NOTED THAT THE IMAGES WERE TAKEN OF THE ¿AS RECEIVED¿ VALVE. THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS 30MMH2O. THE VALVE WAS HYDRATED FOR ABOUT 25 HOURS. THE VALVE WAS VISUALLY INSPECTED: NO DEFECTS WERE NOTED. THE CATHETER WAS VISUALLY INSPECTED AND A CUT WAS NOTED. THE VALVE WAS FLUSHED, THE VALVE PASSED THE TEST NO OCCLUSION WAS NOTED. THE VALVE WAS LEAK TESTED, NO LEAKS WERE NOTED. THE CATHETERS WERE IRRIGATED WITH PURIFIED WATER, NO OCCLUSION WAS NOTED, BUT LEAKED FROM THE CUT IN THE CATHETER. REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE PRODUCT CODE 82-3113, WITH LOT CTPBCW, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK IN 15TH JANUARY 2016. THE ROOT CAUSE TO THE CUT IN THE CATHETER IS PROBABLY DUE TO A SHARP OBJECT COMING INTO CONTACT WITH THE CATHETER. AS NOTED IN THE IFU SILICONE HAS A LOW CUT / TEAR RESISTANCE. NO LEAKAGE WAS FOUND WITH THE VALVE. BASED ON THE RESULTS OF THIS INVESTIGATION, NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME, THIS COMPLAINT IS CLOSED.
(B)(4). UPON COMPLETION OF THE INVESTIGATION A FOLLOW UP REPORT WILL BE FILED.
IT IS SAID THAT THERE IS A HOLE ON THE SILICONE HOUSING OF THE SHUNT DEVICE, AND AS A RESULT, THERE WAS LEAKAGE OF SALINE WHEN THE SURGEON FILLED IT WITH SALINE BEFORE IMPLANTATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 374584 | HAKIM PROGRAMMABLE VALVE MICRO W/RICKHAM | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC. / MEDOS S.A. | CTPBCW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |