FDA Adverse Event Injury Summary report: N

TOUCH ULTRASOUND

MDR report key: 5720420 · Received June 13, 2016

Report

Report Number
1317307-2016-00003
Event Type
Injury
Date Received
June 13, 2016
Date of Event
June 3, 2016
Report Date
July 11, 2016
Manufacturer
CARESTREAM HEALTH INC
Product Code
IYN
PMA / PMN Number
K152467
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CARESTREAM HEALTH INC., NOTIFIED SITES WITH IMPACTED DEVICES ON (B)(6 2016 AND THAT UNTIL A SOFTWARE CORRECTION IS MADE, CUSTOMER SITES SHOULD NOT USE THE 3 DISTANCE GENERIC VOLUME TOOL BUT INSTEAD USE THE TOOLBOX VOLUME CALCULATION FUNCTION, WHICH IS WORKING CORRECTLY,. A SOFTWARE UPDATE IS CURRENTLY BEING TESTED AND VALIDATED AND WILL BE DEPLOYED AS A CORRECTION TO THE FIELD WHEN READY. CARESTREAM HEALTH INC., WILL SUBMIT NOTICE OF A CORRECTION TO THE FDA AS SOON AS POSSIBLE, BUT CERTAINLY BEFORE (B)(6) 2016.

Additional Manufacturer Narrative · 1

CARESTREAM HEALTH HAS INVESTIGATED THIS ADVERSE EVENT INCIDENT AND HAS DETERMINED THIS IS A SOFTWARE DEFECT WHICH REQUIRES A CORRECTION TO THE INSTALL BASE. A REPORT OF CORRECTIONS AND REMOVALS NOTIFICATION WAS SUBMITTED TO THE FDA ON JUNE 24, 2016.

Description of Event or Problem · 1

BASED ON FEEDBACK FROM PRODUCT DEMONSTRATIONS, CARESTREAM HEALTH INC HAS BEEN MADE AWARE OF A SOFTWARE DEFECT WITH INCORRECT VOLUME CALCULATION WHEN USING THE 3 DISTANCE GENERIC VOLUME TOOL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
375827 TOUCH ULTRASOUND TOUCH ULTRASOUND IYN CARESTREAM HEALTH INC

Patients

Seq Age Sex Outcome Treatment
1 Other