TOUCH ULTRASOUND
Report
- Report Number
- 1317307-2016-00003
- Event Type
- Injury
- Date Received
- June 13, 2016
- Date of Event
- June 3, 2016
- Report Date
- July 11, 2016
- Manufacturer
- CARESTREAM HEALTH INC
- Product Code
- IYN
- PMA / PMN Number
- K152467
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
CARESTREAM HEALTH INC., NOTIFIED SITES WITH IMPACTED DEVICES ON (B)(6 2016 AND THAT UNTIL A SOFTWARE CORRECTION IS MADE, CUSTOMER SITES SHOULD NOT USE THE 3 DISTANCE GENERIC VOLUME TOOL BUT INSTEAD USE THE TOOLBOX VOLUME CALCULATION FUNCTION, WHICH IS WORKING CORRECTLY,. A SOFTWARE UPDATE IS CURRENTLY BEING TESTED AND VALIDATED AND WILL BE DEPLOYED AS A CORRECTION TO THE FIELD WHEN READY. CARESTREAM HEALTH INC., WILL SUBMIT NOTICE OF A CORRECTION TO THE FDA AS SOON AS POSSIBLE, BUT CERTAINLY BEFORE (B)(6) 2016.
CARESTREAM HEALTH HAS INVESTIGATED THIS ADVERSE EVENT INCIDENT AND HAS DETERMINED THIS IS A SOFTWARE DEFECT WHICH REQUIRES A CORRECTION TO THE INSTALL BASE. A REPORT OF CORRECTIONS AND REMOVALS NOTIFICATION WAS SUBMITTED TO THE FDA ON JUNE 24, 2016.
BASED ON FEEDBACK FROM PRODUCT DEMONSTRATIONS, CARESTREAM HEALTH INC HAS BEEN MADE AWARE OF A SOFTWARE DEFECT WITH INCORRECT VOLUME CALCULATION WHEN USING THE 3 DISTANCE GENERIC VOLUME TOOL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 375827 | TOUCH ULTRASOUND | TOUCH ULTRASOUND | IYN | CARESTREAM HEALTH INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |