FDA Adverse Event Injury Summary report: N

JYM INFUSION PUMP

MDR report key: 5719790 · Received June 7, 2016

Report

Report Number
MW5062756
Event Type
Injury
Date Received
June 7, 2016
Date of Event
June 8, 2016
Report Date
June 7, 2016
Manufacturer
JIAN YUAN MEDICAL TECHNOLOGY CO. LTD
Product Code
FRN
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

WE WANT TO REPORT PROBLEMS IN MEDICAL INFUSION PUMPS. THE INFUSION PUMP OFTEN USES A DROP SENSOR (DROP COUNTER) TO INCREASE ACCURACY, BUT THIS INFRARED DROP SENSOR CAN BE SPOOFED BY ANOTHER PERSON. THROUGH EXPERIMENTS, WE FOUND OUT THAT THE ATTACKER CAN MANIPULATE THE INFUSED VOLUME OF FLUID WITH HIS OR HER INFRARED SOURCE. THE INFUSED VOLUME CAN BE MANIPULATED FROM 65% TO 330% IN OUR EXPERIMENTS. IT IS BECAUSE THE INFUSION PUMP SENSES DRIPLETS THROUGH THE DROP SENSOR JUST WITH VARIATIONS OF THE OUTPUT OF THE DROP SENSOR. BELOW ARE INFUSION PUMPS THAT WE EXPERIMENTED ON. PRODUCT 1: JSB-1200. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
359688 JYM INFUSION PUMP JYM INFUSION PUMP FRN JIAN YUAN MEDICAL TECHNOLOGY CO. LTD JSB-1200

Patients

Seq Age Sex Outcome Treatment
1 Other