FDA Adverse Event Malfunction Summary report: N

UNKNOWN

MDR report key: 5719716 · Received June 13, 2016

Report

Report Number
3002808486-2016-00498
Event Type
Malfunction
Date Received
June 13, 2016
Date of Event
February 20, 2012
Report Date
November 21, 2024
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
ATTORNEY
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). (B)(4). G1) NAME AND ADDRESS FOR IMPORTER SITE: COOK MEDICAL INCORPORATED (CMI) 400 DANIELS WAY BLOOMINGTON, IN 47404 REGISTRATION NO.: 3005580113. THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE PROVIDED UPON CONCLUSION. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). MANUFACTURER REFERENCE #(B)(4). H11) CORRECTED DATA BASED ON NEW INFORMATION RECEIVED: B1: ADVERSE EVENT TO PRODUCT PROBLEM. H1: SERIOUS INJURY TO MALFUNCTION. IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE VIA THE OPERATIVE NOTE STATING 'DEVICE UNABLE TO BE RETRIEVED, PAIN, DISCOMFORT". COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. FILTER RETRIEVAL IS OCCASIONALLY DIFFICULT. THIS IS WELL-KNOWN FROM PUBLISHED SCIENTIFIC LITERATURE WHERE FILTER RETRIEVALS ARE REFERRED TO AS SIMPLE VS. COMPLEX. SEVERAL CASE REPORTS PUBLISHED IN SCIENTIFIC LITERATURE DESCRIBE COMPLEX CASES WITH SUCCESSFUL ENDOVASCULAR FILTER RETRIEVALS USING ADDITIONAL, ADVANCED TECHNIQUES. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: A1 (PATIENT IDENTIFIER), B5. INVESTIGATION: INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. PER QUALITY ENGINEERING REVIEW, THE ADDITIONAL INFORMATION PROVIDED FOR THIS COMPLAINT DOES NOT CHANGE THE PREVIOUS INVESTIGATION CONCLUSION. THEREFORE, NO NEW INVESTIGATION ACTIVITIES WILL BE CONDUCTED AT THIS TIME. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

THIS ADDITIONAL INFORMATION RECEIVED ON 04/05/2017 AS FOLLOWS: PLAINTIFF ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2012 VIA THE RIGHT COMMON FEMORAL VEIN DUE TO DEEP VEIN THROMBOSIS. PLAINTIFF IS ALLEGING DEVICE IS UNABLE TO BE RETRIEVED, PAIN AND DISCOMFORT.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED AS FOLLOWS: PATIENT IS ALLEGING THE FILTER IS UNABLE TO BE RETRIEVED DUE TO THE HIGH RISK OF POSSIBLE PERFORATION, BLEEDING, NEPHROTOXICITY, AND CONTRAST ALLERGY.

Additional Manufacturer Narrative · 1

MANUFACTURER REF# (B)(4). CATALOG # UNKNOWN AS INFORMATION WAS NOT PROVIDED BUT REFERRED TO AS EITHER GUNTHER TULIP MREYE, GUNTHER TULIP VENA CAVA FILTER, COOK CELECT VENA CAVA FILTER OR COOK CELECT PLATINUM. SINCE CATALOG# IS UNKNOWN THE 510(K) COULD BE EITHER K000855, K032426, K061815, K073374, K090140, K112119, K121057 OR K121629. INVESTIGATION IS STILL IN PROGRESS

Additional Manufacturer Narrative · 1

(B)(4). CATALOG # UNKNOWN BUT REFERRED TO AS A COOK CELECT FILTER. SINCE CATALOG# IS UNKNOWN THE 510(K) COULD BE EITHER K061815, K073374 OR K090140. IT HAS NOT BEEN POSSIBLE TO INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVING WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56.

Description of Event or Problem · 1

DESCRIPTION ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK IVC FILTER". PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Description of Event or Problem · 1

DESCRIPTION ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK IVC FILTER". ADDITIONAL INFORMATION RECEIVED 25FEB2017: CELECT FILTER IMPLANTED ON (B)(6) 2012. PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
376257 UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE

Patients

Seq Age Sex Outcome Treatment
1 31 YR Female Life Threatening