MOSAIQ
Report
- Report Number
- 2950347-2016-00030
- Event Type
- Injury
- Date Received
- June 10, 2016
- Date of Event
- April 8, 2016
- Report Date
- September 19, 2016
- Manufacturer
- IMPAC MEDICAL SYSTEMS, INC.
- Product Code
- IYE
- PMA / PMN Number
- K141572
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
A REVIEW OF SCREEN SHOTS SUPPLIED BY THE CUSTOMER WAS PERFORMED AS PART OF THE INVESTIGATION FOR THIS. IT APPEARS THAT THE USER HAD SELECTED THE WRONG UNITS US VS. SI IN THE VITALS/LAB DATA DIALOG BOX. THIS SELECTION IS DONE BY USING RADIO BUTTONS. THERE MAY HAVE BEEN A MISTREATMENT HOWEVER IT WAS DUE TO USER ERROR. IT SHOULD BE OBVIOUS THAT THE BSA CALCULATED IS NOT CORRECT.. THE MANUFACTURER'S INVESTIGATION IS ON-GOING AND FURTHER INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS COMPLETED.
THE INVESTIGATION WAS COMPLETED BY CONDUCTING A THOROUGH EVALUATION OF THE PRODUCT AND THE REPORTED INFORMATION. MOSAIQ IS WORKING AS DESIGNED AND INTENDED. THE VITALS/LAB DATA WINDOW DEFAULTS TO US STANDARD UNITS FOR WEIGHT AND HEIGHT. THE "US STANDARD" CHOICE IS CLEARLY INDICATED (RADIO BUTTON SELECTED) AND THE UNITS ON WEIGHT (LB) AND HEIGHT (IN) ARE ALSO CLEARLY LABELED. IN SPITE OF THIS, THE USER ENTERED KG AND CM VALUES WHICH RESULTED IN AN INCORRECT DOSE. THEREFORE THIS ISSUE WAS A RESULT OF HUMAN ERROR.
IT WAS REPORTED BY THE CUSTOMER THAT WHEN CALCULATING THE BSA FOR CHEMOTHERAPY MOSAIQ DEFAULTS TO IMPERIAL MEASUREMENTS. THE CUSTOMER WORKS WITH METRIC MEASUREMENTS AND HAS ENTERED THESE VALUES INTO THE IMPERIAL MEASUREMENT FIELDS. BASED ON THE AVAILABLE INFORMATION A MISTREATMENT MAY HAVE OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 372976 | MOSAIQ | ACCELERATOR, LINEAR, MEDICAL | IYE | IMPAC MEDICAL SYSTEMS, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |