FDA Adverse Event Injury Summary report: N

MOSAIQ

MDR report key: 5716644 · Received June 10, 2016

Report

Report Number
2950347-2016-00030
Event Type
Injury
Date Received
June 10, 2016
Date of Event
April 8, 2016
Report Date
September 19, 2016
Manufacturer
IMPAC MEDICAL SYSTEMS, INC.
Product Code
IYE
PMA / PMN Number
K141572
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF SCREEN SHOTS SUPPLIED BY THE CUSTOMER WAS PERFORMED AS PART OF THE INVESTIGATION FOR THIS. IT APPEARS THAT THE USER HAD SELECTED THE WRONG UNITS US VS. SI IN THE VITALS/LAB DATA DIALOG BOX. THIS SELECTION IS DONE BY USING RADIO BUTTONS. THERE MAY HAVE BEEN A MISTREATMENT HOWEVER IT WAS DUE TO USER ERROR. IT SHOULD BE OBVIOUS THAT THE BSA CALCULATED IS NOT CORRECT.. THE MANUFACTURER'S INVESTIGATION IS ON-GOING AND FURTHER INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS COMPLETED.

Additional Manufacturer Narrative · 1

THE INVESTIGATION WAS COMPLETED BY CONDUCTING A THOROUGH EVALUATION OF THE PRODUCT AND THE REPORTED INFORMATION. MOSAIQ IS WORKING AS DESIGNED AND INTENDED. THE VITALS/LAB DATA WINDOW DEFAULTS TO US STANDARD UNITS FOR WEIGHT AND HEIGHT. THE "US STANDARD" CHOICE IS CLEARLY INDICATED (RADIO BUTTON SELECTED) AND THE UNITS ON WEIGHT (LB) AND HEIGHT (IN) ARE ALSO CLEARLY LABELED. IN SPITE OF THIS, THE USER ENTERED KG AND CM VALUES WHICH RESULTED IN AN INCORRECT DOSE. THEREFORE THIS ISSUE WAS A RESULT OF HUMAN ERROR.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT WHEN CALCULATING THE BSA FOR CHEMOTHERAPY MOSAIQ DEFAULTS TO IMPERIAL MEASUREMENTS. THE CUSTOMER WORKS WITH METRIC MEASUREMENTS AND HAS ENTERED THESE VALUES INTO THE IMPERIAL MEASUREMENT FIELDS. BASED ON THE AVAILABLE INFORMATION A MISTREATMENT MAY HAVE OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
372976 MOSAIQ ACCELERATOR, LINEAR, MEDICAL IYE IMPAC MEDICAL SYSTEMS, INC.

Patients

Seq Age Sex Outcome Treatment
1 Other