SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2016-06869
- Event Type
- Malfunction
- Date Received
- June 9, 2016
- Date of Event
- April 30, 2016
- Report Date
- April 30, 2016
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING ASSOCIATED WITH THE RIGHT VENTRICULAR LEAD, THE CRITERIA FOR THE RIGHT VENTRICULAR LEAD INTEGRITY ALERT WERE MET, THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE, AND OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER). ONE NST EPISODE WITH V-V INTERVALS LESS THAN 220 MS ON (B)(4)-2016; 91 V-SICS SINCE LAST SESSION 4.8 DAYS AGO. LEAD FAILURE PREDICTOR TRIGGERED ON (B)(4)-2016 FOR V-SICS AND NSTS. RIGHT VENTRICULAR PACE IMPEDANCE STEADY AT APPROX. 493 OHMS THROUGH (B)(4)-2015, RISES OUT OF RANGE BY (B)(4)-2015.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THERE WAS NOISE ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 363362 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6935M55 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |