VENTED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2016-00253
- Event Type
- Malfunction
- Date Received
- June 8, 2016
- Date of Event
- May 10, 2016
- Report Date
- May 10, 2016
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED A CRACK IN THE CHAMBER DOME BENEATH ONE OF THE PORTS. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR LOT NUMBER 151107. CONCLUSION: THE POSITION AND NATURE OF THE CRACKING UNDER ONE OF THE PORTS INDICATES THAT THE CRACK WAS MOST LIKELY CAUSED BY EXCESSIVE PRESSURE DURING USE. THIS WAS FURTHER CONFIRMED BY INFORMATION ABOUT THE VENTILATOR SETTINGS USED DURING THE INCIDENT. EVERY MR290 CHAMBER IS PRESSURE TESTED FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DUE TO CRACKS AND OTHER CAUSES. ANY CHAMBER WHICH FAILS THIS TEST IS REJECTED. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THAT THE MAXIMUM OPERATING PRESSURE IS 8 KPA. AND ALSO CONTAIN THE FOLLOWING WARNING: USE OF THE MR290 ABOVE THE MAXIMUM OPERATING PRESSURE MAY LEAD TO CRACKING, WATER LEAKAGE AND, ON RARE OCCASIONS, COULD LEAD TO A LOSS OF VENTILATION PRESSURE.
A HOSPITAL IN (B)(6) REPORTED THAT AN MR290V HUMIDIFICATION CHAMBER DOME HAD CRACKED AFTER 20 HOURS OF PATIENT USE. THEY FURTHER REPORTED THAT THIS AFFECTED THE PATIENT THERAPY DUE TO A 50% LEAKAGE OF THE INSPIRATORY VOLUME. NO FURTHER PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 360534 | VENTED HUMIDIFICATION CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 1511070306 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |