CAPTURE-R READY-SCREEN (3)
Report
- Report Number
- 1034569-2016-00130
- Event Type
- Malfunction
- Date Received
- June 8, 2016
- Date of Event
- May 10, 2016
- Report Date
- June 7, 2016
- Manufacturer
- IMMUCOR, INC.
- Product Code
- KSZ
- UDI-DI
- 10888234001454
- PMA / PMN Number
- 102707/0.0
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
IMMUCOR TECHNICAL SUPPORT PERFORMED A REMOTE REVIEW OF THE INSTRUMENT CAMERA IMAGES: BATCH (B)(4) TESTED ON (B)(6) 2016 USING LOTS R732 WITH 221623 NEGATIVE REACTIONS FOR CELLS 1 AND 2 VISUALLY NEGATIVE POSITIVE REACTION FOR CELL 3 AT 2+. THE PI LAB CONFIRMED THE REACTIVITY OF THE E ANTIGEN ON CELL 2 OF RETENTION CAPTURE-R READY-SCREEN (3), LOT R732, ON THE ECHO, USING RETENTION CAPTURE-R READY INDICATOR RED CELL, LOT 221635 AND ANTI-E, LOT BA631. CONTROLS PERFORMED AS EXPECTED AND ALL CELLS RESULTED AS EXPECTED. RETENTION PRODUCT PERFORMED AS EXPECTED. PI LAB PERFORMED A SCREEN ASSAY ON CUSTOMER'S RETURN SAMPLE, SAMPLE (B)(6) (PER CUSTOMER PATIENT HAS AN ANTI-E) ON THE ECHO, USING RETENTION CAPTURE-R READY-SCREEN (3), LOT R732 AND CAPTURE-R READY INDICATOR RED CELL, LOT 221635. CONTROLS PERFORMED AS EXPECTED. CUSTOMER'S SAMPLE RESULTED POSITIVE ON CELL 3 (E-) AND EQUIVOCAL (PIN HOLES) ON CELLS 1 (E-) AND 2 (E+). SAMPLE RETESTED USING THE SAME REAGENTS. CONTROLS PERFORMED AS EXPECTED. CUSTOMER'S SAMPLE RESULTED POSITIVE ON CELL 3 (E-), EQUIVOCAL (VISUALLY NEGATIVE) ON CELL 1 (E-), AND NEGATIVE ON CELL 2 (E+). WE WERE ABLE TO REPRODUCE CUSTOMER'S OBSERVATION WITH SUBMITTED SAMPLE. THIS ISSUE APPEARS TO BE UNIQUE TO THE SAMPLE.
ON (B)(6) 2016, A CUSTOMER REPORTED UNEXPECTED NEGATIVE ANTIBODY SCREENS WHEN USING CAPTURE-R READY-SCREEN (3) (CRRS 3) ON A GALILEO ECHO INSTRUMENT. THE SAMPLE CONTAINS AN ANTI-E.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 361289 | CAPTURE-R READY-SCREEN (3) | REAGENT RED BLOOD CELLS | KSZ | IMMUCOR, INC. | R732 | 10888234001454 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |